beautypg.com

Digitizing oscilloscope modules, Mainframes – Atec Tektronix-TLA700 Series User Manual

Page 4

background image

Digitizing Oscilloscope
Modules

GENERAL

Number of Channels per Module –
TLA 7D2, TLA 7E2: 4 channels.
TLA 7D1, TLA 7E1: 2 channels.
Sample Rate –
TLA 7E1, TLA 7E2: 5 GS/s on all channels.
TLA 7D1, TLA 7D2: 2.5 GS/s on all channels.
Bandwidth (at probe tips) –
TLA 7E1, TLA 7E2:

100 mV to 10 V range: 1 GHz.
50 mV to 99.8 mV range: 900 MHz.
20 mV to 49.8 mV range: 600 MHz.

All others:

500 MHz.
TLA 7D1, TLA 7D2: 500 MHz on all channels in
all ranges.

Memory Depth –
15,000 samples per channel in all modes.
Number of Mainframe Slots Required – 2.

VERTICAL SYSTEM

Input Sensitivity Range – 10 mV to 100 V full
scale.
Vertical Resolution – 8 bits (256 levels).
DC Gain Accuracy – ±1.5% of full scale range.
Analog Bandwidth Selections – 20 MHz,
250 MHz, and Full.
Input Coupling – AC, DC, or GND.
Input Impedance Selections – 1 M

in paral-

lel with 10 pF, or 50

.

AC Coupled Lower Frequency Limit –

10 Hz

when AC 1 M

coupled,

200 kHz when AC

50

coupled.

Maximum Input Voltage at Probe
Connector –
300 VRMS, but no greater than
±420 V peak (1 M

or ground input coupling).

PROBE INPUT CHARACTERISTICS

Probe Input Interface – TEKPROBE

probe

interface.
Input Loading – Less than 1 pF in parallel with
1 M

with either P6243 or P6245.

Usable Input Voltage Range at Probe Tip -
P6243 Probe: ±8 V. P6245 Probe: ±18 V.

ACQUISITION SYSTEM

Sample Rate Range – 200 ps to 200 ms in 1,
2.5, 5 sequence.
Timebase Accuracy – ±100 ppm over any inter-
val

1 ms.

Record Length Range – 512 to 15,000 samples
per channel in all modes.
Acquisition Modes – Single-shot, repetitive.

TRIGGER SYSTEM

Trigger Modes – Normal, auto.
Trigger Position – Anywhere in the acquired
record (pre-fill can be set anywhere from 0% to
100%).
Trigger Types – Edge, pulse width, timeout,
glitch, runt, slew rate, logic pattern, setup-and-
hold violation.
Trigger Actions – Trigger, trigger all, set signal,
arm, immediate, wait for system trigger.
Edge Trigger – Conventional level driven trigger,
positive or negative slope, on any channel or
external trigger input. Coupling Selections: DC,
AC, noise reject, HF reject, LF reject.
Pulse Width Trigger – Trigger on width of posi-
tive or negative pulse, either within or not within
selectable time limits; settable from 2 ns to 1 s.
Timeout Trigger – Triggers when a pulse fails
to complete when specified; settable from 2 ns
to 1 s.
Glitch Trigger – Trigger on (or reject) glitches of
positive, negative, or either polarity; settable from
2 ns to 1 s. Minimum glitch width: 2.0 ns, with
200 ps resolution (2 ns to 10 ns settings).
Runt Pulse Trigger – Trigger on a pulse that
crosses one threshold but fails to cross a second
threshold before crossing the first again; settable
from 2 ns to 1 s.
Slew Rate Trigger – Trigger on pulse edge rates
that are either faster or slower than a set rate,
edges can be rising, falling, or either; settable
from 2 ns to 1 s.
Logic Pattern Trigger – Triggers when a logical
combination (AND, OR) of all the input channels
(Hi, Lo, Don’t Care) stays true or false for a speci-
fied period of time; settable from 2 ns to 1 s.
Setup-and-hold Trigger – Trigger on violations
of both setup time and hold time between clock
and data which are on separate input channels;
setup time settable from –100 ns to +100 ns in
200 ps increments; hold time settable from –1 ns
to +102 ns; minimum settable window of setup
time + hold time is 2 ns.

Digitizing Oscilloscope Modules
P h y s i c a l C h a r a c t e r i s t i c s

Dimensions

mm

in.

Height

262

10.3

Width

61

2.4

Depth

381

15

Weight

kg

lb.

Net

2.7

5.8

Shipping

5.8

12.8

P6243 Probe Cable Length – 1.3 m (51 in.).
P6245 Probe Cable Length – 1.3 m (51 in.).
Other Compatible Probes – P6139A, P6204,
P5100, P6205, P6408, P6563A, P6701B, P6703B,
AM 503S, TCP202, P6247, P6248, P6249.

Mainframes

GENERAL (TLA 714, TLA 720, TLA 7XM)

Instrument Slots –
TLA 714: Holds 4 single-wide or 2 double-wide
modules.
TLA 720: Holds 10 single-wide or 5 double-wide
modules.
TLA 7XM: Holds 12 single-wide or 6 double-wide
modules.

PC CHARACTERISTICS (TLA 714 AND
TLA 720)

Operating System – Microsoft Windows.
Processor – Pentium

®

.

DRAM –
TLA 714: 64 MB SDRAM (128 MB with Opt. 1S).
TLA 720: 128 MB SDRAM.
Display Memory – 2 MB.
Sound – Built-in PC speaker transducer; multi-
media sound can be added via PC Card interface.
Replaceable Hard Disk Drive –
TLA 714: 6 GB (12 GB with Opt. 1s).
TLA 720: 12 GB.
A second internal fixed hard disk is available for
both mainframes with TLA 7UP Opt. 04.
CD Rom – 24x, Internal.
Floppy Disk Drive – Built-in 3.5 in. 1.44 MB
drive.
Year 2000 Compliant – Yes.

INTEGRAL CONTROLS (TLA 714 ONLY)

Front-Panel Display – Size: 10.4 in. diagonal.
Type: Active-matrix color TFT LCD with backlight.
Resolution: 800 x 600.
Colors: 16.8 M (true color).
Simultaneous Display Capability – Both the
front-panel and external displays can be used
simultaneously at 800 x 600 resolution.
Front-panel Knobs – Special function knobs for
instrument control.
Front-panel Qwerty Keypad – Mini-QWERTY
keypad and Hex keypad.
Front-panel Pointing Device – Glidepoint fin-
ger pad.

EXTERNAL PERIPHERAL INTERFACES
(TLA 714 AND TLA 720)

External Display Port Type – Female DB15
SVGA connector.
External Display Resolution – Up to 1600 x
1200 non-interlaced @ 256 colors.
External Display Compatibility – DDC2B
(Dynamic Display Configuration 2).
External Keyboard Port Type – PS2 mini-DIN.
External Mouse Port Type – PS2 mini-DIN.
Parallel Interface Port Type – IEEE 1284-C
connector (comes standard with adapter to
female DB25 connector).

L O G I C A N A L Y Z E R S • T L A 7 0 0 S E R I E S

4

Characteristics