82-253290 E01 March 2008
9
TEST SYSTEM FOR CONDUCTED AND
RADIATED IMMUNITY
NSG 4070
Application for IEC/EN 61000-4-6, calibration set-up with current injection probe
Power recommendation, achievable test levels with 6 dB attenuator, 0.5 dB cable loss,
max. insertion loss of the coupling device and AM with 80% modulation depth
Application for IEC/EN 61000-4-6, EUT set-up with EM clamp or current injection probe
and for example with use of a monitoring probe
AE-side
EUT-side
EM clamp
Insulating
Ground plane
6 dB attenuator
Auxiliary
equipment
150 kHz ... 230 MHz (1000 MHz)
EM-Koppelzange
KEMZ 801
EUT
RF out
Amp in
< +10 dBm
Amp out
ch.1 < +27 dBm
ch.2 < +20 dBm
ch.3 < +20 dBm
Power meter
NSG 4070
Power
USB
Tuning
StSize
StSize
StSize
Local
Back
Stop
Run
Hold
0
.
1
4
7
8
5
2
3
kHz
dBm
6
9
MHz
dBµV
Hz
V
Enter
Step
1
STO
FRQ
LVL
RCL
Step
2
Step
3
MOD
2nd
Help
RF
ON/OFF
Compact generator NSG 4070
with built-in power amplifier
EUT
AE
Equipment
under test
Monitoring probe
Power meter ch. 1
Insulating
Calibration
jig
Current injection
probe
Power meter ch. 1
Ground plane
RF out
Amp in
< +10 dBm
Amp out
ch.1 < +27 dBm
ch.2 < +20 dBm
ch.3 < +20 dBm
Power meter
NSG 4070
Power
USB
Tuning
StSize
StSize
StSize
Local
Back
Stop
Run
Hold
0
.
1
4
7
8
5
2
3
kHz
dBm
6
9
MHz
dBµV
Hz
V
Enter
Step
1
STO
FRQ
LVL
RCL
Step
2
Step
3
MOD
2nd
Help
RF
ON/OFF
Compact generator NSG 4070
with built-in power amplifier
150 Ω
termination
150/50 Ω
adapter
6 dB
attenuator
Amplifi er module:
20 W
30 W
75 W
CDN:
15 V EMF
18 V EMF
30 V EMF
EM clamp (KEMZ 801):
8 V EMF
11 V EMF
17 V EMF
Current injection clamp (CIP 9136):
5 V EMF
6 V EMF
10 V EMF (typ.)