NSG 4070 datasheet
4
TEST SYSTEM FOR CONDUCTED AND
RADIATED IMMUNITY
NSG 4070
Test and measurement routines
Firmware: Generator mode
Firmware: Immunity mode
Sweep:
frequency sweep, level sweep
Modulation:
AM, AM PC (peak conservation), pulse modulation and external
Others:
free parameter setting from 9 kHz to 1 GHz, high power mode
using power amplifi er
Level:
constant or slope test levels, max test levels depending on power
amplifi er, test routine for IEC 61000-4-6 level 1 to 3 and X up to
30 V EMF, for BCI tests levels in mA or dBµA
Test methods IEC 61000-4-6:
CDN, EM clamp, current clamp and direct injection, clamp injection
with test level control using monitoring probe
Test methods BCI:
substitution method with optional use of the monitoring probe,
closed loop method with power limitation (factor adjustable)
Sweep:
frequency sweep, sweep function linear, steps per decade, percen-
tal and as requested in ISO 11452
Modulation:
AM, AM PC (peak conservation), pulse modulation, external or
mixed (e.g. 1 kHz AM internal modulated with 1 Hz PM external)
EUT monitoring:
Individual confi guration of the ports, function to check or prepare
the EUT monitoring, display of EUT monitoring results during the
test, in the result fi le and in the test report
Calibration:
Test set-up and monitoring probe calibration, display, store
and recall function of calibration fi les (limitation of fi le numbers only
by the disk space, typical >340 fi les)
EUT threshold search:
test interrupt for manual or automatic change of frequency or stress
level
Store and recall:
store and recall function of test confi gurations, calibration results
and test results (number of fi les is only limited by the disk space,
typical >340 fi les), supports USB sticks
Component check:
quick check of system components, e.g. cable, attenuator max.
52 dB/ 54 dB/ 58 dB attenuation for 20 W/ 30 W/ 75 W amplifi er,
max. +16 dB gain at 27 dBm output level
Additional features:
free parameter setting from 9 kHz to 1 GHz, supports external
power amplifi er, directional coupler and attenuator
Firmware: Main generator menu
Firmware: Immunity test setup
Firmware: Calibration result