beautypg.com

Test system for conducted and radiated immunity, Nsg 4070 – Atec Teseq-Schaffner-NSG4070-75 User Manual

Page 2

background image

NSG 4070 datasheet

2

TEST SYSTEM FOR CONDUCTED AND

RADIATED IMMUNITY

NSG 4070

Technical specifications

Generator

RF
Frequency range:

9 kHz – 1 GHz

Resolution:

1 Hz

Reference frequency:

10 MHz Reference output


RF Level
Level range:

-60 dBm to +10 dBm

Resolution:

0.1 dB

Settling time:

10 ms


Amplitude modulation
Modulation depth:

0 – 100%

Modulation frequency range: 1 Hz – 50 kHz
Frequency resolution:

1 Hz


Pulse modulation
Rise / fall time (10% / 90%):

< 1 µs

Modulation frequency range: 1 Hz – 50 kHz
Frequency resolution:

1 Hz

Duty cycle:

10% to 90%


External modulation
Delay time:

< 1 µs / 180°

Period:

min. 20 µs

Pulse width:

min. 10 µs

Power meter

Frequency range:

9 kHz – 1 GHz

Linear measurement range

channel 1:

-15 dBm to +27 dBm

channel 2,3:

-25 dBm to +20 dBm

Max. input/no damage

channel 1-3:

+28 dBm

Noise level:

>6 dB below the measurement range

Input return loss:

>20 dB (below 500 MHz), >17 dB (500 MHz to 1 GHz)

Connector:

BNC socket, 50 Ω