Atec Rohde-Schwarz-SMA100A User Manual
Page 3
![background image](/manuals/319549/3/background.png)
All-purpose instrument
❙Frequency range of 9 kHz to 3 GHz (R&S®SMA-B103/
R&S®SMA-B103L) or 6 GHz (R&S®SMA-B106/
R&S®SMA-B106L)
❙Frequency, level and LF sweeps
❙Phase-continuous frequency setting
❙AM, broadband FM/φM (R&S®SMA-B20 or
R&S®SMA-B22), pulse modulation
❙Built-in LF generator up to 1 MHz, optional multifunction
generator (R&S®SMA-K24) up to 10 MHz
❙Optional low-jitter clock synthesizer up to 1.5 GHz
(R&S®SMA-B29)
❙Power measurement using R&S®NRP-Zxx power sensors
❙Optional power analysis (R&S®SMA-K28) using
R&S®NRP-Zxx power sensors for scalar network analysis
or automatic pulse parameter measurement
▷
page 6
Aerospace and defense applications
❙Pulse modulator with excellent characteristics
(on/off ratio of > 100 dB (meas.) for f < 5.5 GHz,
rise/fall time typ. < 7 ns)
❙Pulse generator integrated as standard
❙Optional high-performance pulse generator with
minimum pulse width of 10 ns (R&S®SMA-K23)
❙Optional generation of versatile pulse sequences/pulse
trains (R&S®SMA-K27)
❙Optional chirp modulation (R&S®SMA-B20 or
R&S®SMA-B22)
❙Optional VOR/ILS modulation (R&S®SMA-K25)
❙Optional DME modulation/analysis (R&S®SMA-K26)
❙Optional operating altitude up to 4600 m (R&S®SMA-B46)
❙Optional removable mass storage (CompactFlash™ card,
R&S®SMA-B80)
▷
page 8
Intuitive operating concept and versatile
interfaces
❙Color display with 480 × 272 pixel
❙Intuitive user interface with graphical display of signal
flow (block diagram)
❙Context-sensitive online help
❙Remote control via GPIB, LAN or USB
❙Selectable control language (SCPI or remote control
emulation of various signal generators)
❙Remote operation by browser or VNC client
❙USB connectors (e.g. for keyboard, mouse, memory
stick)
❙Support of R&S®NRP-Zxx power sensors for precise
power measurements
▷
page 11
Excellent signal quality
❙Very low SSB phase noise of typ. –134 dBc
(20 kHz carrier offset, f = 1 GHz, 1 Hz measurement
bandwidth), typ. –139.5 dBc with the R&S®SMA-B22
enhanced phase noise performance option
❙Wideband noise of –162 dBc (meas.) with carrier offset
> 40 MHz, f = 1 GHz, level = 9 dBm, 1 Hz measurement
bandwidth
❙Nonharmonics < –96 dBc (carrier offset > 10 kHz,
f < 750 MHz, with the R&S®SMA-B22 option)
❙High-stability reference oscillator as standard
❙Very low phase noise at low frequencies due to internal
division of the fundamental frequency range (750 MHz to
1500 MHz) down to 6.6 MHz
▷
page 4
Ideal for use in production
❙Very short frequency/level setting times of
< 1.5 ms/1 ms (meas.) across the entire frequency and
level range, < 450 µs in List mode
❙Fast Hopping mode with flexibly addressable frequency
and level pairs, as fast as normal List mode
❙Frequency setting time of typ. 10 µs within a bandwidth
of up to 80 MHz due to direct access to the DDS-based
synthesizer (with the R&S®SMA-B20 or R&S®SMA-B22
option; FM EXTERNAL DIGITAL mode)
❙Very high level accuracy and repeatability
❙High output power of up to +18 dBm, overrange up to
+28 dBm
❙Electronic attenuator with built-in overvoltage protection
over entire frequency range
❙Minimum space requirements due to compact size
(only two height units)
▷
page 5
Rohde & Schwarz R&S®SMA100A Signal Generator 3
R&S®SMA100A
SignalGenerator
Benefitsand
keyfeatures