beautypg.com

Atec Rohde-Schwarz-SMP Series User Manual

Page 12

background image

Microwave Signal Generator SMP

12

Internal modulation generator

Frequency

0.4/1/3/15 kHz ± 3%

Open-circuit voltage at LF connector

1 V ± 1% (R

out

= 10

Ω, R

load

>200

Ω)

(peak value)

LF generator option SM-B2

Waveforms

sinewave, triangular, squarewave,
noise

Frequency range

Sinewave, noise
Triangular, squarewave

0.1 Hz to 500 kHz
0.1 Hz to 50 kHz

Resolution

0.1 Hz

Frequency accuracy

<1 x 10

–4

Frequency response (sinewave)

up to 100 kHz
up to 500 kHz

<0.3 dB
<0.5 dB

Distortion (20 Hz to 100 kHz)

<0.1% (for level >0.5 V)

Open-circuit voltage at LF connector

Resolution
Setting accuracy at 1 kHz

1 mV to 4 V (R

out

= 10

Ω, R

load

>200

Ω)

1 mV
±1% + 1 mV

Frequency setting time
(after IEC/IEEE-bus delimiter)

<10 ms

Pulse generator option SMP-B14

Operating modes

single pulse, delayed pulse, double
pulse

Active trigger edge

positive or negative

Pulse repetition period

Resolution
Accuracy

100 ns to 85 s
5 digits, min. 20 ns
same as reference frequency

Pulse width

Resolution
Accuracy

40 ns to 1 s
4 digits, min. 20 ns
<(5% of reading ± 3 ns)

Pulse delay

Resolution
Accuracy

40 ns to 1 s
4 digits, min. 20 ns
±5% of reading
–10 ns to +20 ns

Double pulse

Resolution
Accuracy

60 ns to 1 s
4 digits, min. 20 ns
±5% of reading
–10 ns to +20 ns

Trigger delay

<50 ns

PULSE modulation input

Input level
Input impedance

TTL (HCT)
50

Ω or 10 kΩ

SYNC output

TTL level (HC), 40 ns pulse width

VIDEO output

TTL level (HC)

PP

PD

PW

SYNC output

VIDEO output

RF output

The pulse generator option enables the pulse delay PD, pulse width PW and
pulse repetition period PP to be set with high accuracy and resolution.

RF control output

Frequency range

2 GHz to 20 GHz

Level

approx. 0 dBm

Sweep

digital sweep in discrete steps

RF sweep, AF sweep

Operating modes

Sweep range

Step width linear
Step width logarithmic

AF sweep with option SM-B2
automatic, single-shot, manual or ex-
ternally triggered, linear or logarithmic

user-selectable
0.01% to 50%

Level sweep

Operating modes

Sweep range
Step width

automatic, single-shot, manual or
externally triggered, logarithmic
0.1 dB to 20 dB
0.1 dB to 20 dB

Step time

Resolution

10 ms to 1 s
0.1 ms

Markers

3, user-selectable

MARKER output

TTL/HC logic signal, selectable polarity

X output

0 V to 10 V

BLANK output

TTL/HC logic signal, selectable polarity

TRIGGER input

TTL/HCT logic signal, polarity of active
trigger edge selectable

STOP input

TTL/HCT logic signal, selectable
polarity

LIST mode

frequency and level values can be
stored and read out fast; permissible
level variation range: 20 dB

Operating modes

automatic, single-shot, manual or
externally triggered

Max. length of list

2003 pairs of frequency and level
values

Step time

Resolution

(1 ms to 1 s) + the less of 5 ms/GHz or
50 ms
0.1 ms

Memory for instrument settings

Storable settings

50

Memory sequence modes

Operating modes

automatic, single-shot, manual or
externally triggered

Step time

Resolution

50 ms to 60 s
1 ms

Auxiliary interface with option SMP-B18

V/GHz output

output voltage proportional to frequen-
cy, 0.5 V/GHz

9)

or 1 V/GHz selectable

Z output

user-selectable level range
–10 V to +10 V