Atec Picotest-G5100A User Manual
Page 4
Specifications
Display
Graph mode for visual verification of signal settings
Capability
Standard
waveforms
Sine, Square, Ramp, Triangle, Pulse, Noise, DC
Built-in Arbitrary
Exponential Rise and Fall, Negative Ramp, Sin(x)/x,
Cardiac
Waveform Characteristics
Sine
Frequency
1 µHz to 50 MHz
Amplitude
Flatness
1,2
(Relative to 1
KHz)
0.1dB(<100 KHz)
0.15dB(<5 MHz)
0.3dB(<20 MHz)
0.5dB(<50 MHz)
Harmonic
Distortion
2,3
(unit: dBc)
DC to 20 KHz
-70 (< 1Vpp) -70 (
≥ 1Vpp)
20 KHz to 100 KHz
-65 (< 1Vpp) -60 (
≥ 1Vpp)
100 KHz to 1 MHz
-50 (< 1Vpp) -45 (
≥ 1Vpp)
1 MHz to 20 MHz
-40 (< 1Vpp) -35(
≥ 1Vpp)
20 MHz to 50 MHz
-35 (< 1Vpp) -30(
≥ 1Vpp)
Total Harmonic
Distortion
2,3
DC to 20 KHz, Output
≥ 0.5
Vpp
THD+N
≤ 0.06%
Spurious
2,4
(non-harmonic)
DC to 1 MHz
-70 dBc
1 MHz to 50 MHz
-70 dBc + 6 dB/Octave
Phase Noise
(10K Offset)
-115/dBC/Hz, Typical
When F
≥ 1 MHz, V ≥ 0.1
Vpp
Square
Frequency
1 µHz to 25 MHz
Rise/Fall Time
< 10ns
Overshoot
< 2%
Variable Duty
Cycle
20% to 80% (to 10 MHz)
40% to 60% (to 25 MHz)
Asymmetry
1% of period + 5ns (@ 50% duty)