beautypg.com

Specifications, General, Ambient conditions – Atec Fluke-1760 User Manual

Page 6

background image

6 Fluke Corporation Fluke 1760 Three-Phase Power Quality Recorder Topas

Specifications

General

Warranty

years

Quality system

Developed and manufactured as per ISO 9001: 000

Reference conditions

Environment temp.: 3 °C ± °C (74 °F ± °F)
Power supply: 30 V ± 10 %
Power frequency: 50 Hz/60 Hz
Signal: declared input voltage U

din

Averaging: 10 minute intervals

Intrinsic uncertainty

All specified intrinsic uncertainties are valid under these reference conditions

Display

The Fluke 1760 features LED indicators for the status of the 8 channels, phase sequence,

power supply (mains or battery), memory usage, time synchronization, and data transfer.

Power LED

Permanent light: normal power supply from mains
OFF: supply via internal battery in case of a power failure

Channel LEDs

3-color LEDs per channel for:

• overload condition

• under load condition signal level in nominal range

Data memory

GB flash memory depending on model

Memory model

Linear

Interfaces

Ethernet (100 MB/s), RS 3, external modem via RS 3

Baud rate for RS 232

9600 Baud to 115 kBaud

Dimensions (H x W x D)

35 mm x 300 mm x 65 mm (13 in x 11.8 in x .6 in)

Weight (without accessories)

appr. 4.9 kg (10.8 lb)

Calibration interval

1 year recommended for Class-A, otherwise years

Ambient conditions

Working temperature range

-0 °C to 50 °C (-4 °F to 1 °F)

Storage temperature range

-0 °C to 60 °C (-4 °F to 140 °F)

Reference temperature

3 °C ± °C (74 °F ± °F)

Climatic class

B (IEC 654-1), -0 °C to 50 °C (-4 °F to +1 °F)

Max. operating altitude

000 m: max. 600 V CAT IV*, power supply: 300 V CAT III

5000 m: max 600 V CAT III*, power supply: 300 V CAT II

*depending on sensor

Housing

insulated, robust plastics housing

EMC

Emission

Class-A as per IEC/EN 6136-1

Immunity

IEC/EN 6136-1