5 Fluke Corporation Fluke 1760 Three-Phase Power Quality Recorder Topas

Overview of measurement functions

Statistical

evaluation

Power quality statistics according to EN50160 and DISDIP tables

like ITIC, CEBEMA, ANSI

EN 50160

Event list

Dips, swells, and interruptions are detected and stored in the

event list. Also any trigger which fires generates an event,

which is added to this list. Together with the event, rms values,

transients, and fast transients can be stored.
The event list shows the exact time of occurrence, as well as

duration and magnitude. Several parameters of the events can

be sorted. The user can thereafter select an event for further root

cause analysis.

Event List

DISDIP

Continuous
recording

The Fluke 1760 logs rms values, together with corresponding
minimum and maximum values, continuously – with the
following time aggregations:
1 Day
10 min
Free interval, e.g. 15 min, h

Voltage

Current

Power P, Q, S

Power Factor

kWh

Flicker

Unbalance

Frequency

Harmonics/Interharmonics

Triggered

recordings

RMS with adjustable aggregation time between 10 ms
(1/2 cycle), 20ms (1 cycle), 200 ms (10/12 cycles) or 3 sec
(150/180 cycles).

Calculating rms values, harmonics, and interharmonics is
performed synchronous to the power frequency.

Basic aggregation for harmonics and interharmonics is 200 ms

Oscilloscope: Sample rate is 10.24 kHz for all 8 channels

Fast Transients: Sample rate is selectable from 100 kHz to
10 MHz for channel 1-4

Oscilloscope

Fast Transients

FFT of Fast Transients

Mains

signaling

Voltage and current on all phases and neutral

Online mode

Variable refresh rate. This feature allows verification of set up of
the instrument and delivers a quick overview.

Oscilloscope
Transients
Events