Atec LeCroy-WAVEPRO-950-960 User Manual
Page 4
Horizontal zoom expand to 2 pts/div, magnify to 50000X
Auto Scroll automatically scans and displays any zoom or math trace.
Rapid Signal Processing
Processor
PowerPC
Processing Memory
Up to 256 Mbytes
Realtime Clock
Dates, hours, minutes, seconds
Internal Waveform Memory
Waveform
M1, M2, M3, M4 (Store full-length waveforms with 16 bits/data point)
Zoom and Math
Four traces A, B, C, D with chained trace capability
Setup Storage
Front Panel and Instrument
Status
Four non-volatile memories and floppy drive are standard. Hard drive and memory card are optional.
CustomDSO
Customize and access scope settings with up to 6 CustomDSO files stored in non-volatile Virtual Disk
(VDisk).
Interface
Remote Control
Full control of all front panel controls and internal functions via RS232C, GPIB, or Ethernet
RS-232-C
Asynchronous transfer rate of up to 115.2 kbaud
GPIB Port
Full control via IEEE – 4888.2; configurable as talker/listener for computer control and data transfer
Ethernet (optional)
10 BaseT Ethernet interface
Floppy Drive
Internal, DOS-format, 3.5" high-density
PC Card Slot (optional)
Supports memory and hard drive cards
External Monitor Port
Standard
15-pin D-Type VGA-compatible
Centronics Port
Parallel printer interface
Internal Graphics Printer
(optional)
Hard copy output in <10 seconds or strip chart mode up to 200 cm/div
Pass/Fail and Trigger Output Front panel Cal BNC output provides choice of Cal Signal, Pass/Fail Condition, Trigger Ready, or Trigger
Out signals
Outputs
Calibrator Signal
500 Hz – 2 MHz square wave or 25 ns pulse; 0.05 to +1.0 Volt into 1 MOhms output on front panel
BNC
Control Signals
Trigger ready, trigger out, pass/fail status.