Atec Flexim-F610 User Manual
Page 7
TSFLUXUS_F601V1-5-1US_Lus, 2014-02-25
7
Technical Specification
FLUXUS® F601
communication
interface
RS232/USB
serial data kit
software (all Windows™ ver-
sions)
- FluxData: download of measurement data, graphical presentation,
conversion to other formats (e.g. for Excel™)
- FluxKoef: creating medium data sets
- FluxSubstanceLoader: upload of medium data sets
сable
RS232
adapter
RS232 - USB
transport case
dimensions
19.7 x 15.7 x 7.5 in
outputs
The outputs are galvanically isolated from the transmitter.
number
see standard scope of supply on page 9, max. on request
accessories
output adapter (if number of outputs > 4)
current output
range
0/4 to 20 mA
accuracy
0.1 % of reading ±15 μA
active output
R
ext
< 200 Ω
passive output
U
ext
= 4 to 16 V, depending on R
ext
R
ext
< 500 Ω
frequency output
range
0 to 5 kHz
open collector
24 V/4 mA
binary output
optorelay
26 V/100 mA
binary output as alarm output
- functions
limit, change of flow direction or error
binary output as pulse output
- pulse value
0.01 to 1000 units
- pulse width
1 to 1000 ms
inputs
The inputs are galvanically isolated from the transmitter.
number
see standard scope of supply on page 9, max. 4
accessories
input adapter (if number of inputs > 2)
temperature input
type
Pt100/Pt1000
connection
4-wire
range
-238 to +1040 °F
resolution
0.01 K
accuracy
±0.01 % of reading ±0.03 K
current input
accuracy
0.1 % of reading ±10 μA
passive input
R
i
= 50 Ω, P
i
< 0.3 W
- range
-20 to +20 mA
voltage input
range
0 to 1 V
accuracy
0.1 % of reading ±1 mV
internal resistance
R
i
= 1 MΩ
FLUXUS
F601