Atec Flexim-F610 User Manual

Page 7

background image

TSFLUXUS_F601V1-5-1US_Lus, 2014-02-25

7

Technical Specification

FLUXUS® F601

communication
interface

RS232/USB

serial data kit
software (all Windows™ ver-
sions)

- FluxData: download of measurement data, graphical presentation,

conversion to other formats (e.g. for Excel™)

- FluxKoef: creating medium data sets
- FluxSubstanceLoader: upload of medium data sets

сable

RS232

adapter

RS232 - USB

transport case
dimensions

19.7 x 15.7 x 7.5 in

outputs

The outputs are galvanically isolated from the transmitter.

number

see standard scope of supply on page 9, max. on request

accessories

output adapter (if number of outputs > 4)
current output

range

0/4 to 20 mA

accuracy

0.1 % of reading ±15 μA

active output

R

ext

< 200 Ω

passive output

U

ext

= 4 to 16 V, depending on R

ext

R

ext

< 500 Ω

frequency output

range

0 to 5 kHz

open collector

24 V/4 mA
binary output

optorelay

26 V/100 mA

binary output as alarm output
- functions

limit, change of flow direction or error

binary output as pulse output
- pulse value

0.01 to 1000 units

- pulse width

1 to 1000 ms

inputs

The inputs are galvanically isolated from the transmitter.

number

see standard scope of supply on page 9, max. 4

accessories

input adapter (if number of inputs > 2)
temperature input

type

Pt100/Pt1000

connection

4-wire

range

-238 to +1040 °F

resolution

0.01 K

accuracy

±0.01 % of reading ±0.03 K
current input

accuracy

0.1 % of reading ±10 μA

passive input

R

i

= 50 Ω, P

i

< 0.3 W

- range

-20 to +20 mA
voltage input

range

0 to 1 V

accuracy

0.1 % of reading ±1 mV

internal resistance

R

i

= 1 MΩ

FLUXUS

F601