Flow transmitter, Technical data – Atec Flexim-F610 User Manual
Page 6
![background image](/manuals/319265/6/background.png)
6
TSFLUXUS_F601V1-5-1US_Lus, 2014-02-25
FLUXUS® F601
Technical Specification
Flow Transmitter
Technical Data
FLUXUS
F601
design
portable
measurement
measurement principle
transit time difference correlation principle,
automatic NoiseTrek selection for measurements with high gaseous or solid content
flow velocity
0.03 to 82 ft/s
repeatability
0.15 % of reading ±0.03 ft/s
medium
all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference
principle)
temperature compensation
corresponding to the recommendations in ANSI/ASME MFC-5.1-2011
accuracy
1
with standard calibration
with advanced calibration
(optional)
with field calibration
2
±0.5 % of reading ±0.03 ft/s
flow transmitter
power supply
100 to 240 V/50 to 60 Hz (power supply unit),
10.5 to 15 V DC (socket at transmitter),
integrated battery
battery
Li-Ion, 7.2 V/4.5 Ah
operating time (without outputs, inputs and backlight): > 14 h
power consumption
< 6 W
number of flow measuring
channels
2
signal attenuation
0 to 100 s, adjustable
measuring cycle (1 channel) 100 to 1000 Hz
response time
1 s (1 channel), option: 70 ms
housing material
PA, TPE, AutoTex, stainless steel
degree of protection
NEMA 4
dimensions
see dimensional drawing
weight
4.2 lb
fixation
QuickFix pipe mounting fixture
ambient temperature
14 to 140 °F
display
2 x 16 characters, dot matrix, backlight
menu language
English, German, French, Dutch, Spanish
measuring functions
physical quantities
volumetric flow rate, mass flow rate, flow velocity,
heat flow (if temperature inputs are installed)
totalizer
volume, mass, optional: heat quantity
calculation functions
average, difference, sum
diagnostic functions
sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times
data logger
loggable values
all physical quantities, totalized values and diagnostic values
capacity
> 100 000 measured values
1
for transit time difference principle, reference conditions and v > 0.49 ft/s
2
reference uncertainty < 0.2 %
±1.6 % of reading ±0.03 ft/s
±1.2 % of reading ±0.03 ft/s