Wall thickness measurement (optional), Technical data – Atec Flexim-F610 User Manual
Page 24
24
TSFLUXUS_F601V1-5-1US_Lus, 2014-02-25
FLUXUS® F601
Technical Specification
Wall Thickness Measurement (optional)
The pipe wall thickness is an important pipe parameter which has to be determined exactly for a good mea-
surement. However, the pipe wall thickness often is unknown.
The wall thickness probe can be connected to the transmitter instead of the flow transducers and the wall
thickness measurement mode is activated automatically.
Acoustic coupling compound is applied to the wall thickness probe which then is placed firmly on the pipe.
The wall thickness is displayed and can be stored directly in the transmitter.
Technical Data
Cable
technical type
DWR1NZ7
measuring range
1
in
0.04 to 9.8
resolution
in
0.0004
accuracy
1 % ± 0.004 in
medium temperature
°F
-4 to +392,
short-time peak max. 932
сable
type
2616
length
ft
4
1
The measuring range depends on the attenuation of the ultrasonic signal in the
pipe. For strongly attenuating plastics (e.g. PFA, PTFE, PP) the measuring range is smaller.
type
2616
ambient temperature
°F
<392
cable jacket
material
FEP
outer diameter
in
0.2
color
black
shield
x
DWR1NZ7