Page 7

background image

Portaflow 330 Technical Datasheet

Portaflow 330 Technical Datasheet

7

(Issue 2.1)

Technical data

Analogue output:

Pulse output:

Range –

4–20mA
0–20mA
0–16mA

Output type –

One open collector opto-isolated digital
output

Resolution –

0.1% of full scale

Pulse repetition –

Up to 500 pulses/sec (depending on
pulse width)

Alarm current –

Adjustable between 0–26mA

Pulse width –

500ms for 1pulse/s
5ms for 100 pulses/s

Isolation –

1500V Opto-isolated

Max current –

150mA

Maximum load – 620 Ohms

General specification:

DSP measurement technique –

The PF330 uses ‘transit time’ measurement with a sampling resolution of 50 pico-
second, with a continuous signal level indication to the display

Flow velocity range (bi-directional) –

Minimum velocity 0.1m/s
Max velocity 20m/s

Flow measurement accuracy –

±0.5% to ±2% of flow reading for flow rate >0.2m/s and Pipe ID >75mm
±3% of flow reading for flow rate >0.2m/s and Pipe ID in range 13mm - 75mm
±6% of flow reading for flow rate < 0.2m/s

Flow velocity corrected for Reynolds number over entire velocity range

Repeatability –

±0.5% of measured value or ±0.02m/s whichever is the greater

Response time –

< 500ms depending on pipe diameter

Selectable flow units –

VELOCITY: m/sec, ft/sec

VOLUME: l/s, l/min, l/h, gal/min, gal/h, USgals/min, USgals/h, Barrel/h,

Barrel/day, m³/s, m³/min, m³/h

Selectable volume units –

l, gal, USgals, Barrel, m³

Total volume –

12 Digits - forward and reverse

Display languages –

The following operator languages can be selected from the PF330 set-up menu

English
French
German
Swedish

Italian
Spanish
Portuguese

Russian
Norwegian
Dutch