KeyTek ECAT
®
Model E505A
Plug-in module that produces the lightning surge
waveforms required by FCC Part 68
ELECTRICAL
Waveforms
<10/>160µs, 50-1650V ±10%; peak short-circuit
current is 200A, -0% +10% when the open-circuit
voltage is set to 1500V
<10/>560µs, 50-880V ±10%; peak short-circuit
current is 100A, -0% +10% when the open-circuit
voltage is set to 800V
<2/>10µs, 100-2750V ±10%; peak short-circuit
current is 1000A, -0% +10% when the open-circuit
voltage is set to 2500V
Note: All voltage and current specifications are
minimum values in accordance with FCC Part 68
Surge Repetition Rate
1 shot/18 seconds for all waves except <2/>10µs
which is 1 shot/24 seconds
Line Sync Accuracy
±15º with optional coupler/decoupler
Minimum System Requirements
E100 series control center
Options
E505A-VI - adds voltage and current monitoring
KeyTek ECAT
®
Model E506-4W
Plug-in module to produce the 2/10µs surges required by
Telcordia GR-1089 CORE for up to five-wire (four terminal)
testing
ELECTRICAL
Waveforms
<2/>10µs, 50-800V, 100A/terminal with 800V open-
circuit voltage
<2/>10µs, 50-1500V, 100A/terminal with 1500V
open-circuit voltage
<2/>10µs, 100-2500V, 500A/terminal with 2500V
open-circuit voltage
<2/>10µs, 200-5000V, 500A/terminal with 5000V
open-circuit voltage
Tolerances
All peak open-circuit voltages and short-circuit
currents are -0%/+20%
Outputs
Front panel terminals for connection to T1, R1, T2,
R2 and Ground
Surge
Repetition Rate: 1 shot/16 seconds
Minimum System Requirements
E100 series control center
Options
E506-4W-VI - adds voltage and current monitoring
KeyTek ECAT
®
Model E508 and E508-12P
Plug-in modules to produce the 10/360µs surges required by Telcordia GR-1089 CORE
E508 WAVEFORMS
Open-Circuit Voltage
10/360µs, 50-1100V, -0/+15% in 1 volt steps.
Tip and ring outputs independent and isolated to
ensure true, three-terminal simultaneous testing of
up to 12 pair. Waveforms as defined by Telcordia
GR-1089-CORE
Short-Circuit Current
100A/side -0/+15% at a voltage setting of 1.0kV
Front Time Tolerance
-30/+0% for voltage and current
Duration Tolerance
-0/+30%
Surge Repetition Rate
1 shot/50 seconds
E508-12P WAVEFORMS
Open-Circuit Voltage
10/360µs, 50-1100V, -0/+15% in 1 volt steps.
Tip and ring outputs independent and isolated to
insure true, three-terminal simultaneous testing of
up to 12 pair. Waveforms as defined by Telcordia
GR-1089-CORE.
Short-Circuit Current
25A/side -0/+15% at a voltage setting of 1.0kV
Front Time Tolerance
-25%/+0 for voltage; -30%/+0 for current
Duration Tolerance
-0/+30%
Surge Repetition Rate
1 shot/150 seconds
Minimum System Requirements
E100 series control center
Options
E508-VI - adds voltage and current monitoring