Ess-2000, Specifications – Atec Noiseken-ESS-2000 User Manual
Page 5
Electrostatic Discharge Simulator
4
ESS-2000
■
SPECIFICATIONS
Parameters
ESS-2000 specifications
Output voltage
0.20 ~ 30.0kV ±5%
Polarity
Positive or negative
Charging resistance
10MΩ(53MΩ for combination with TC-815R discharge gun)
Discharge mode
Air discharge and contact discharge
Operation
mode
IEC
severity
level
Level setting
1, 2, 3, 4
Discharge interval
0.05 ~ 600.0 s
No. of times of discharge
1 ~ 60000 times
Manual
Discharge interval
0.05 ~ 600.0 s
No. of times of discharge
1 ~ 60000 times
Setting storage function
Up to 10 conditions storable
Sweep
Starting voltage
±0.20 ~ 30.0 kV
Ending voltage
±0.20 ~ 30.0 kV
Step voltage
0.00 ~ 30.0 kV
Discharge interval
0.05 ~ 600.0 s
No.of times of discharge
1 ~ 60000 times
Setting storage function
Up to 10 conditions storable
Program
Voltage setting
±0.20 ~ 30.0 kV
Discharge interval
0.05 ~ 600.0 s
No. of times of discharge
1 ~ 60000 times
No. of steps
30 steps maximum
No. of programs
Up to 50 conditions storable
No. of program units
Up to 100 conditions can be set.
Display element
LCD with back light
Character display
English or Japanese
Setting method
Ten-key pad, Rotary control, Function keys
Auxiliary function
Upper limit voltage setting function / Trigger switch select function
Auto stop function / Inversion on the screen function
Memory function
Contents of each setting and last operation display are backed up for more than 3 months with battery
full charged.
External interface functions
GP-IP connecting I/F / Warning light connecting I/F
External trigger input I/F / Elimination probe connecting I/F
Printer interface
Conforming to simple CENTRONIX I/F
Contents of print
Currently applied voltage/ Contents of various settings / Current temperature and humidity(option)
Power supply
100 ~ 240 VAC 50/60 Hz
Operating temperature and humidity
15 ~ 35°C 25 ~ 75% (No dewing shall occur.)
Dimensions and weight
(W)250 × (H)324 × (D)320 mm Approx. 8.0 kg