dc VOLTaGE OUTpUT (2 Years, 23°c ± 5°c)
oUtPUt Voltage: 0 to +15V DC.
oUtPUt aCCUraCy: ±(0.05% + 3mV).
ProgrammIng resolUtIon: 1mV.
reaDbaCK aCCUraCy
1
: ±(0.05% + 3mV).
reaDbaCK resolUtIon: 1mV.
oUtPUt Voltage settlIng tIme: 5ms to within stated accuracy.
loaD regUlatIon: 0.01% + 2mV.
lIne regUlatIon: 0.5mV.
stabIlIty
2
: 0.01% + 0.5mV.
measUrement tIme CHoICes: 0.01 to 10PLC
7
, in 0.01PLC steps.
aVerage reaDIngs: 1 to 10.
reaDIng tIme
1, 8, 9
: 31ms, typical.
transIent resPonse:
High Bandwidth
Low Bandwidth
transient recovery time
13
<40µs
3
or <60µs
4
<80µs
3
or <100µs
4
transient Voltage Drop
<75mV
3
or <100mV
4
<250mV
3
or <400mV
4
remote sense: 1V max. drop in each lead. Add 2mV to the voltage load regulation specification for
each 1V change in the negative output lead due to load current change. Remote sense required.
Integrity of connection continually monitored. If compromised, output will turn off automatically
once settable window (±0 to ±8V) around normal voltage exceeded.
VaRIaBLE OUTpUT IMpEdaNcE
range: 0 to 1.00
Ω in 0.01Ω steps. Value can be changed with output on.
dc cURRENT (2 Years, 23°c ± 5°c)
oUtPUt CUrrent (2302):
0–4V: 5A max.
>4V: I
MAX
= 60W/(V
OUT
+6) (not intended to be operated in parallel).
ContInUoUs aVerage oUtPUt CUrrent (2306, 2306-PJ):
Channel #2 (Charger) oFF:
I = 50W/(V
SET
channel 1 + 6V); 5A max.
Channel #2 (Charger) on:
I = (50W – Power consumed by channel #2)/(V
SET
channel 1 + 6V); 5A max.
The power consumed by channel #2 is calculated as:
Channel #2 sourcing current:
Power consumed = (V
SET
channel 2 + 6V)
× (current supplied)
Channel #2 sinking current:
Power consumed = 5
× (sink current)
Peak currents can be a maximum of 5A provided the average current is within the above limits.
ContInUoUs aVerage sInK CUrrent:
Channel #2 (Charger) oFF:
0–5V: 3A max.
5–15V: Derate 0.2A per volt above 5V. Compliance setting controls sinking.
Channel #2 (Charger) on:
Available current = (50W – Power consumed by channel #2)/5; 3A max. (0–5V).
Derate 0.2A per volt above 5V.
soUrCe ComPlIanCe aCCUraCy: ±(0.16% + 5mA)
5
.
ProgrammeD soUrCe ComPlIanCe resolUtIon: 1.25mA.
reaDbaCK aCCUraCy
1
: 5a range:
±(0.2% + 200µA).
5ma range: ±(0.2% + 1µA) (2302 and 2306).
500ma range: ±(0.2% + 20µA) (2306-PJ only).
reaDbaCK resolUtIon: 5a range:
100µA.
5ma range: 0.1µA (2302 and 2306).
500ma range: 10µA (2306-PJ only).
loaD regUlatIon: 0.01% + 1mA.
lIne regUlatIon: 0.5mA.
stabIlIty
4
: 0.01% + 50µA.
measUrement tIme CHoICes: 0.01 to 10PLC
7
, in 0.01PLC steps.
aVerage reaDIngs: 1 to 10.
reaDIng tIme
1, 8, 9
: 31ms, typical.
pULSE cURRENT MEaSUREMENT OpERaTION
trIgger leVel:
5a CUrrent range
5a range:
5mA to 5A, in 5mA steps.
1a range:
1mA to 1A, in 1mA steps.
100ma range: 0.1mA to 100mA, in 100µA steps.
500ma CUrrent range (2306-PJ)
500ma range: 0.5mA to 500mA, in 0.5mA steps.
100ma range: 0.1mA to 100mA, in 100µA steps.
10ma range: 100µA to 10mA, in 100µA steps.
trIgger Delay: 0 to 100ms, in 10µs steps.
Internal trIgger Delay: 15µs.
HIgH/low/aVerage moDe:
measurement aperture settings: 33.3µs to 833ms, in 33.3µs steps.
average readings: 1 to 100.
PUlse CUrrent measUrement aCCUraCy
11
(2 years, 23°C ±5°C):
aperture
accuracy
±(% reading + offset + rms noise
10
)
5a Range
500ma Range (2306-pJ)
<100 µs
0.2% + 900 µA + 2 mA
0.2% + 90 µA + 2 mA
100 µs – 200 µs
0.2% + 900 µA + 1.5 mA
0.2% + 90 µA + 1.5 mA
200 µs – 500 µs
0.2% + 900 µA + 1 mA
0.2% + 90 µA + 1 mA
500 µs – <1 PLC
0.2% + 600 µA + 0.8 mA
0.2% + 60 µA + 0.8 mA
1 PLC
12
0.2% + 400 µA + 0 mA
0.2% + 40 µA + 0 mA
>1 PLC
0.2% + 400 µA + 100 µA
0.2% + 40 µA + 100 µA
BURST MOdE cURRENT MEaSUREMENT
measUrement aPertUre: 33.3µs.
ConVersIon rate: 3650/second, typical.
Internal trIgger Delay: 15µs.
nUmber oF samPles: 1 to 5000.
transFer samPles aCross Ieee bUs In bInary moDe: 4800 bytes/s, typical.
LONG INTEGRaTION MOdE cURRENT MEaSUREMENT
2302, 2306: Available on 5A range only.
2306-PJ: Available on 5A and 500mA current ranges.
measUrement tIme
6
: 850ms (840ms) to 60 seconds in 1ms steps.
dIGITaL VOLTMETER INpUT (2 Years, 23°c ± 5°c)
InPUt Voltage range: –5 to +30V DC.
InPUt ImPeDanCe: 2M
Ω typical.
maxImUm Voltage (either input terminal) wItH resPeCt to oUtPUt low: –5V, +30V.
reaDIng aCCUraCy
1
: ±(0.05% + 3mV).
reaDIng resolUtIon: 1mV.
ConneCtor: HI and LO input pair part of Output #1’s terminal block.
measUrement tIme CHoICes: 0.01 to 10PLC
7
, in 0.01PLC steps.
aVerage reaDIngs: 1 to 10.
reaDIng tIme
1, 8, 9
: 31ms, typical.