Atec Fluke-2638A User Manual
Page 8
8 Fluke Corporation Fluke 2638A Hydra Series III Data Acquisition System/Digital Multimeter
AC voltage
AC voltage specifications are for ac sine wave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range.
Maximum input
Front panel: 300 V rms or 425 V peak; Rear inputs: 150 V CAT II; 250 V peak
Measurement method
AC-coupled true-rms. Measures the ac component of input with up to 300 V dc bias on any range
AC filter
bandwidth
Slow
20 Hz
Fast
200 Hz
Maximum crest factor
5:1 at full scale
AC voltage input characteristics
Range
Resolution
Measurement rate/resolution
Input impedance
Fast
4.5 digits
Medium
5.5 digits
Slow
6.5 digits
100 mV
100.0000 mV
10 µV
1 µV
0.1 µV
1 MΩ ±2 % shunted
by 150 pF
1 V
1.000000 V
100 µV
10 µV
1 µV
10 V
10.00000 V
1 mV
100 µV
10 µV
100 V
100.0000 V
10 mV
1 mV
100 µV
300 V
300.000 V
100 mV
10 mV
1 mV
AC voltage accuracy
Accuracy is given as ± (% measurement + % of range).
Range
Frequency
24 hour (23 ±1 °C)
90 days (23 ±5 °C)
1 year (23 ±5 °C)
T.C./°C outside
18 °C to 28 °C
100 mV
20 Hz to 20 kHz
0.1 + 0.05
0.11 + 0.05
0.11 + 0.05
0.01 + 0.005
20 Hz to 50 kHz
0.2 + 0.05
0.22 + 0.05
0.22 + 0.05
0.01 + 0.005
50 Hz to 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.05 + 0.01
1 V
20 Hz to 20 kHz
0.1 + 0.05
0.11 + 0.05
0.11 + 0.05
0.01 + 0.005
20 Hz to 50 kHz
0.2 + 0.05
0.22 + 0.05
0.22 + 0.05
0.01 + 0.005
50 Hz to 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.05 + 0.01
10 V
20 Hz to 20k Hz
0.1 + 0.05
0.11 + 0.05
0.11 + 0.05
0.01 + 0.005
20 Hz to 50 kHz
0.2 + 0.05
0.22 + 0.05
0.22 + 0.05
0.01 + 0.005
50 Hz to 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.05 + 0.01
100 V
20 Hz to 20 kHz
0.1 + 0.05
0.11 + 0.05
0.11 + 0.05
0.01 + 0.005
20 Hz to 50 kHz
0.2 + 0.05
0.22 + 0.05
0.22 + 0.05
0.01 + 0.005
50 Hz to 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.05 + 0.01
300 V
20 Hz to 20 kHz
0.1 + 0.05
0.11 + 0.05
0.11 + 0.05
0.01 + 0.005
20 Hz to 50 kHz
0.2 + 0.05
0.22 + 0.05
0.22 + 0.05
0.01 + 0.005
50 Hz to 100 kHz
0.55 + 0.27
0.6 + 0.27
0.6 + 0.27
0.05 + 0.03
DC current
Input protection
0.15 A /600 V resettable PTC
Common mode rejection
140 db at 50 Hz or 60 Hz ±0.1 % (1 kΩ imbalance)
Normal mode rejection
60 db for NPLC of 1 or greater and power line frequency ±0.1 %
DC current input characteristics
Range
Resolution
Measurement rate/resolution
Shunt resistance
(ohms)
Burden voltage
Fast
4.5 digits
Medium
5.5 digits
Slow
6.5 digits
100 µA
100.0000 µA
10 nA
1 nA
0.1 nA
1k Ω
<1 mV
1 mA
1.000000 mA
100 nA
10 nA
1 nA
1k Ω
<1 mV
10 mA
10.00000 mA
1 µA
100 nA
10 nA
10 Ω
<1 mV
100 mA
100.0000 mA
10 µA
1 µA
100 nA
10 Ω
<1 mV