beautypg.com

General specifications – Atec Fluke-2638A User Manual

Page 10

background image

10 Fluke Corporation Fluke 2638A Hydra Series III Data Acquisition System/Digital Multimeter

Resistance input characteristics

Range

Resolution

Measurement Rate / Resolution

Source current

Fast

4.5 digits

Medium

5.5 digits

Slow

6.5 digits

100 Ω

100.0000 Ω

10 mΩ

1 mΩ

0.1 mΩ

1 mA / 4 V

1 kΩ

1.000000 kΩ

100 mΩ

10 mΩ

1 mΩ

1 mA / 4 V

10 kΩ

10.00000 kΩ

1 Ω

100 mΩ

10 mΩ

100 µA / 6 V

100 kΩ

100.0000 kΩ

10 Ω

1 Ω

100 mΩ

100 µA / 12 V

1 MΩ

1.000000 MΩ

100 Ω

10 Ω

1 Ω

10 µA / 12 V

10 MΩ

10.00000 MΩ

1 kΩ

100 Ω

10 Ω

1 µA / 12 V

100 MΩ

100.0000 MΩ

10 kΩ

1 kΩ

100 Ω

0.1 µA / 12 V

Resistance accuracy

Accuracy is given as ± (% measurement + % of range).

Range

24 hour (23 ±1 °C)

90 days (23 ±5 °C)

1 year (23 ±5 °C)

T.C./ °C outside 18 °C

to 28 °C

100 Ω

0.003 + 0.003

0.008 + 0.004

0.01 + 0.004

0.0006 + 0.0005

1 kΩ

0.002 + 0.0005

0.008 + 0.001

0.01 + 0.001

0.0006 + 0.0001

10 kΩ

0.002 + 0.0005

0.008 + 0.001

0.01 + 0.001

0.0006 + 0.0001

100 kΩ

0.002 + 0.0005

0.008 + 0.001

0.01 + 0.001

0.0006 + 0.0001

1 MΩ

0.002 + 0.001

0.008 + 0.001

0.01 + 0.001

0.001 + 0.0002

10 MΩ

0.015 + 0.001

0.02 + 0.001

0.04 + 0.001

0.003 + 0.0004

100 MΩ

0.3 + 0.01

0.8 + 0.01

0.8 + 0.01

0.15 + 0.002

RTD temperature accuracy (4 wire)

Temperature

Accuracy

Fast sample rate (NPLC < 10)

T.C./ °C outside 18 °C to 28 °C

-200 °C

0.015 °C

add 0.02 °C

0.0021 °C

0 °C

0.02 °C

add 0.02 °C

0.0023 °C

300 °C

0.04 °C

add 0.02 °C

0.0028 °C

600 °C

0.06 °C

add 0.02 °C

0.0033 °C

Thermistor temperature accuracy (4 wire)

Temperature

Accuracy

2.2 kΩ thermistor

Accuracy

5 kΩ thermistor

Accuracy

10 kΩ thermistor

-40 °C

0.002 °C

0.003 °C

0.003 °C

0 °C

0.004 °C

0.003 °C

0.003 °C

25 °C

0.01 °C

0.005 °C

0.005 °C

50 °C

0.012 °C

0.007 °C

0.009 °C

100 °C

0.08 °C

0.035 °C

0.017 °C

150 °C

0.35 °C

0.15 °C

0.06 °C

Frequency

Range

Frequency

24 hour

(23 ±1 °C)

90 days

(23 ±5 °C)

1 year

(23 ±5 °C)

T.C./ °C outside

18 °C to 28 °C

100 mV to 300 V

20 Hz to 40 Hz

0.03

0.03

0.03

0.001

40 Hz to 1 MHz

0.006

0.01

0.01

0.001

General specifications

Power

Voltage:

100 V to 264 V in 4 power model selections

(100 V, 120 V, 220 V, 240 V)

Frequency:

47 Hz to 440 Hz

(automatically sensed when turned on)

Power consumption: 36 V A peak (24 W average)

Communication

LAN:

10/100 Mb TCPIP DHC

Remote commands: SCPI command set

Internal memory

53,000 scans

1,000 setup files

File security

Dual level, Administrator, four secure users, Guest status

Web server

Embedded web server—Display of active data channels and alarms

(page release 2013 Q4)

Alarm outputs

6 TTL alarm outputs—assignable to one or multiple channels

Alarm setpoints

2 assignable alarm setpoints/channel , Hi Lo, Hi Hi, Lo Lo

Language menus

English, Chinese, French, German, Japanese, Korean, Portuguese,

Russian and Spanish

Environment

Temperature

Operating:

0 °C to 50 °C

Storage:

20 °C to 70 °C

Warm-up:

One hour to full uncertainty specifications

Relative humidity (non-condensing)

Operating:

0 °C to 28 °C <90 %

28 °C to 40 °C <75 %

40 °C to 50 °C <50 %

Storage:

20 °C to 70 °C<95 %

Physical

Weight:

6 kg (13.2 lbs)

Dimensions:

150 mm x 285 mm x 385 mm

(5.91 in x 11.22 in x 15.16 in

Standard rack height: 3 U