Specifications, Measurement specifications – Atec Fluke-2638A User Manual

Page 7

background image

7 Fluke Corporation Fluke 2638A Hydra Series III Data Acquisition System/Digital Multimeter

Measurement specifications

Accuracy specifications generally are valid for 6.5 digit

resolution mode (unless otherwise noted) after a mini-

mum of one-hour warm-up, and within an operating

temperature range of 18 °C to 28 °C. Derating applies

with a fast sample rate and for operating tempera-

tures outside the limited range. Twenty-four-hour

specifications are relative to calibration standards and

assume a controlled electromagnetic environment per

EN 61326. The confidence level for accuracy specifica-

tions is 95 % within one year of calibration (unless

otherwise noted).

Scan rate

Fast

45 channels per second max (0.03 seconds per channel, depending on function and range)

Medium

16 channels per second (0.1 seconds per channel)

Slow

2 channels per second (0.5 seconds per channel)

Custom

Selectable from 0.2 NPLC to 200 NPLC

Display resolution

4.5 to 6.5 digits, depending on function and sample rate

Specifications

DC voltage

Maximum input

300 V

Common mode rejection

140 db at 50 Hz or 60 Hz ±0.1 % (1k Ω imbalance)

Normal mode rejection

60 db for NPLC of 1 or greater and power-line frequency ±0.1 %

Measurement method

Multi-ramp A/D

A/D linearity

2 ppm of measurement + 1 ppm of range

Input bias current

30 pA at 25 °C

DC voltage input characteristics

Range

Resolution

Measurement rate/Resolution

Input impedance

Fast

4.5 digits

Medium

5.5 digits

Slow

6.5 digits

100 mV

100.0000 mV

10 µV

1 µV

0.1 µV

10 MΩ or >10 GΩ

[1]

1 V

1.000000 V

100 µV

10 µV

1 µV

10 MΩ or >10 GΩ

[1]

10 V

10.00000 V

1 mV

100 µV

10 µV

10 MΩ or >10 GΩ

[1]

100 V

100.0000 V

10 mV

1 mV

100 µV

10 MΩ ±1%

300 V

300.000 V

100 mV

10 mV

1 mV

10 MΩ ±1%

[1] 10 MΩ is default input impedance.

DC voltage accuracy

Accuracy is given as ± (% measurement + % of range).

Range

24 hour (23 ±1 °C)

90 days (23 ±5 °C)

1 year (23 ±5 °C)

T.C./ °C outside

18 °C to 28 °C

100 mV

0.0025 + 0.003

0.0025 + 0.0035

0.0037 + 0.0035

0.0005 + 0.0005

1 V

0.0018 + 0.0006

0.0018 + 0.0007

0.0025 + 0.0007

0.0005 + 0.0001

10 V

0.0013 + 0.0004

0.0018 + 0.0005

0.0024 + 0.0005

0.0005 + 0.0001

100 V

0.0018 + 0.0006

0.0027 + 0.0006

0.0038 + 0.0006

0.0005 + 0.0001

300 V

0.0018 + 0.002

0.0031 + 0.002

0.0041 + 0.002

0.0005 + 0.0003