LEVEL RECORDER MODE

Measurement principle

Selective level measurement at a fixed frequency setting.
Peak

Detection

Root mean square value (RMS), RMS
(integration time = 480 ms, observation time selectable from 480 ms up to 30 min)

Filter

Type

Steep cutoff channel filter

Resolution bandwidth RBW (-6 dB)

40 kHz to 32 MHz (10 steps per decade)

Video bandwidth (VBW)

4 Hz to 32 MHz (depending on the selected RBW)

Measurement range setting (MR)

Set individually from a list or using the “MR Search” function for determining the optimum
measurement range at a given time

Result Type

Peak ACT: Displays the current (actual) value
Peak MAX: Max hold function
RMS ACT: Averaging over a defined time period (0.48 seconds to 30 min)
RMS MAX: Max hold function over the averaged values – with RMS detector only.

Axis

Measurement in the direction of the X, Y, and Z axis
(separate measurement in one direction using an isotropic / three-axis measuring antenna)

Time Averaging

Selectable from 0.96 seconds up to 30 minutes
(0.96 s; 1.2 s; 2.4 s; 3.6 s; 6 s;12 s; 18 s; 30 s;
1 min; 2 min; 3 min; 5 min; 6 min; 10 min;15 min; 20 min; 30 min)

Noise suppression

Identifies whether measured values are above the device noise floor by setting a threshold
(selectable at 0, 3, 6, 10, 15, or 20 dB relative to device noise floor).
Measurement values below the threshold are shown as the absolute threshold value
marked with “<” (less than threshold). Only applies to the numerical result display (Value)

SCOPE MODE (OPTION)

Measurement principle

Selective level measurement at a fixed frequency

Filter

Type

Steep cutoff channel filter

Time Span

500 ns to 24 h

Time Resolution

Selectable from31,25 ns up to 90 min
(0.96 s; 1.2 s; 2.4 s; 3.6 s; 6 s;12 s; 18 s; 30 s;
1 min; 2 min; 3 min; 5 min; 6 min; 10 min;15 min; 20 min; 30 min)

Resolution bandwidth RBW (-6 dB)

40 kHz to 32 MHz (10 steps per decade)

Measurement range setting (MR Range)

Set individually from a list or using the “MR Search” function for determining the optimum
measurement range at a given time

Video bandwidth (VBW)

4 Hz to 32 MHz (depending on the selected RBW)
ACT: Displays the current (actual) value.
Standard: Displays the selected safety standard.
or

Result Type
Depends on detector

MAX: Maximum value within the time resolution interval (corresponds to peak detector).
AVG: Average value within the time resolution interval (corresponds to RMS detector).
MIN: Minimum value within the time resolution interval.
Standard: Displays the selected safety standard.