Atec Narda-SRM-3006 User Manual
Page 14
Single-axis antenna (H-field) 3551/02
Frequency range
9 kHz to 300 MHz
The correction factors determined individually during calibration are stored in an EEPROM and
are applied automatically when used in conjunction with the SRM basic unit.
Antenna type
H-field
Sensor type
Single axis active magnetic loop
Dynamic range
a
0.4 µA/m to 71 mA/m
CW damage level
> 2.65 A/m above 1 MHz
Intrinsic noise display in conjunction with
the SRM basic unit
b
0.12 µA/m for each frequency > 10 MHz with 1 kHz resolution bandwidth RBW
Measurement range limit
(for single CW signal)
b
100 mA/m
RF connector
N connector, 50 Ω
UNCERTAINTY
Frequency range
Single-axis measurement
Extended measurement uncertainty
c
(in conjunction with SRM basic unit and
1.5 m cable)
0.1 - 300 MHz
2.0 dB
Calibration uncertainty
< 1.2 dB
GENERAL SPECIFICATION
Operating temperature range
-10 °C to 50 °C (same as SRM basic unit)
Storage
1K3 (IEC 60721-3) extended to -10 °C to +50 °C
Transport
2K4 (IEC 60721-3)
Climatic
Operating
7K2 (IEC 60721-3)
Storage
1M3 (IEC 60721-3)
Transport
2M3 (IEC 60721-3)
Mechanical
Operating
7M3 (IEC 60721-3)
ESD and EMC
EN 61326:2006
Safety
EN 61010-1:2004
Compliance
EU Guidelines
2003/11/EG 06.02.2003 (PBDE and OBDE)
2002/95/EG 27.01.2003 (RoHS)
2002/96/EG 27.01.2003 (WEEE)
CE (European Community)
Yes
Air humidity
< 29 g/m³ (< 93 % at +30 °C)
Weight 450
g
Dimensions
460 mm length; 43 x 100 mm antenna head dimension
Calibration
183 reference points
The SRM interpolates between reference points
Recommended calibration interval
24 months
Country of origin
Germany
a Typical measurement dynamic range for 10 dB signal to noise radio (RBW = 1 kHz); for frequencies > 10 MHz
b Typical values
c Typical value, with k = 2 (k = extrapolation or correction factor for determining the assessment value); +15 °C to +30 °C
NSTS 1109- ME-E0272C-0.00
14 / 16