Input specifications
Channel 1 & 2 (53131A, 53132A)
1
Channel 1 (53181A)
Frequency range
dc coupled
dc to 225 MHz
ac coupled
1 MHz to 225 MHz (50 Ω)
30 Hz to 225 MHz (1 MΩ)
FM tolerance 25%
Voltage range and sensitivity (Sinusoid)
2
dc to 100 MHz 20 mVrms to ±5 V ac + dc
100 MHz
30 mVrms to ±5 V ac + dc
to 200 MHz
200 MHz
40 mVrms to ±5 V ac + dc
to 225 MHz
(all specified at 75 mVrms
with opt. rear connectors)
3
Voltage range and sensitivity
(Single-shot pulse)
2
4.5 ns to 10 ns 100 mVpp to 10 Vpp
pulse width
(150 mVpp with optional
rear
connectors)
3
>10 ns
50 mVpp to 10 Vpp
pulse width
(100 mVpp with optional
rear
connectors)
3
Trigger level
2
Range
± 5.125 V
Accuracy
± (15 mV + 1% of trigger level)
Resolution 5
mV
Damage level
50 Ω
5 Vrms
0 to 3.5 kHz, 350 Vdc + ac pk
1 MΩ
3.5 kHz to
350 Vdc + ac pk linearly
100 kHz, 1 MΩ derated to 5 Vrms
>100 kHz,
5 Vrms
1 MΩ
Input characteristics
Channel 1 & 2 (53131A, 53132A)
1
Channel 1 (53181A)
Impedance
1 MΩ or 50 Ω
1 MΩ
30 pF
capacitance
Coupling
ac or dc
Low-pass filter 100 kHz, switchable
-20 dB at > 1 MHz
Input
Selectable between Low,
sensitivity Medium,
or
High
(default).
Low is approximately 2x
High
Sensitivity.
Trigger slope Positive or negative
Auto trigger level
Range
0 to 100% in 10% steps
Frequency
> 100 Hz
Input amplitude > 100 mVpp
(No amplitude modulation)
Attenuator
Voltage range x10
Trigger range x10
Input Specifications
4
Channel 3 (53131A, 53132A)
Channel 2 (53181A)
Frequency range
Option 015
100 MHz to 1.5 GHz
(for 53181A
(see Opt. 030 for
only) additional
specs)
Option 030
100 MHz to 3 GHz
Option 050
200 MHz to 5 GHz
Option 124
200 MHz to 12.4 GHz
Power range and sensitivity (Sinusoid)
Option 030
100 MHz to 2.7 GHz:
-27 dBm to +19 dBm
2.7 GHz to 3 GHz:
-21 dBm to +13 dBm
Option 050
200 MHz to 5 GHz:
-23 dBm to +13 dBm
Option 124
200 MHz to 12.4 GHz
-23 dBm to +13 dBm
Damage level
Option 030
5 Vrms
Option 050
+25 dBm
Option 124
+25 dBm
Characteristics
Impedance 50
Ω
Coupling AC
VSWR <
2.5:1
External arm input specifications
5
Signal input range
TTL compatible
Timing Restrictions
Pulse width
> 50 ns
Transition time < 250 ns
Start-to-stop time > 50 ns
Damage level 10
Vrms
External arm input characteristics
5
Impedance 1
kΩ
Input capacitance 17 pF
Start/stop slope Positive or negative
External time base input specifications
Voltage range
200 mVrms to 10 Vrms
Damage level
10 Vrms
Frequency
1 MHz, 5 MHz, and 10 MHz
(53132A 10 MHz only)
Time base output specifications
Output frequency 10 MHz
Voltage
> 1 Vpp into 50 Ω
(centered around 0 V)
1. Specifications and characteristics for Channels
1 and 2 are identical for both common and
separate configurations.
2. Values shown are for X1 attenuator setting.
Multiply all values by 10 (nominal) when using
the X10 attenuator setting.
3. When the 53131A or 53132A are ordered with
the optional rear terminals (Opt. 060), the channel
1 and 2 inputs are active on both front and rear
of the counter. When the 53181A is ordered with
the optional rear terminal, the channel 1 input is
active on both front and rear of the counter. For
this condition, specifications indicated for the rear
connections also apply to the front connections.
4. When optional additional channels are ordered
with Opt. 060, refer to configuration table for
Opt. 060 under ordering info on page 8. There is
no degradation in specifications for this input,
as applicable.
5. Available for all measurements except peak volts.
External arm is referred to as external gate for
some measurements.
4
Instrument Inputs