Available statistics

Mean, Minimum, Maximum, Standard Deviation

Number of measurements 2 to 1,000,000.

Statistics may be collected on all measurements
or on only those which are between the limit
bands. When the limits function is used in
conjunction with statistics, N (number of
measurements) refers to the number of in-limit
measurements. In general, measurement
resolution will improve in proportion to N,
up to the numerical processing limits of
the instrument.

Measurements

Statistics may be collected for all measurements
except peak volts and totalize.

Save and recall

Up to 20 complete instrument setups may
be saved and recalled later. These setups
are retained when power is removed from
the counter.

Rack dimensions (HxWxD)

88.5 mm x 212.6 mm x 348.3 mm

Weight

3.5 kg maximum

Warranty

1 year

Power supply

100 to 120 VAC ± 10% -50, 60 or
400 Hz ± 10% 220 to 240 VAC
± 10% -50 or 60 Hz ± 10%

ac Line selection

Automatic

Power requirements

170 VA maximum (30 W typical)

Environment

0°C to 55°C operating
–40°C to 71°C storage

Remote Interface

GPIB (IEEE 488.1-1987,
IEEE 488.2-1987)

Remote programming language

SCPI-1992.0 (Standard Commands
for Programmable Instruments)

Safety

Designed in compliance with IEC-1010,
UL-3111-1 (draft), CAN/CSA 1010.1

EMC

CISPR-11, EN50082-1,
IEC 801-2, -3, -4

Radiated immunity testing

When the product is operated at maximum
sensitivity (20 mVrms) and tested at 3 V/m
according to IEC 801-3, external 100 to 200 MHz
electric fields may cause frequency miscounts.

Measurement
Statistics

General