Frequency specifications, continued, Sweep time, Resolution bandwidth – Atec Agilent-8564E User Manual
Page 3: Noise sidebands, Residual fm
![background image](/manuals/318705/3/background.png)
3
Sweep time
Range
Span = 0 Hz
50 µs to 6000 s
Span ±100 Hz
RBW
≥
300 Hz
50 ms to 2000 s
RBW
≤
100 Hz
50 ms to 100 ks
Accuracy (Span = 0 Hz)
Sweep time
≥
30 ms
±1% (digitized trace data)
Sweep time <30 ms
±10% (analog trace data)
(non-Option 007)
Sweep time <30 ms
±0.1% (digitized trace data)
(Option 007
1
)
Sweep trigger
delayed, free run, single,
line, video, external
Resolution bandwidth
Range (–3 dB)
1 Hz to 1 MHz in a 1, 3, 10 sequence
and 2 MHz (3 MHz at –6 dB)
Option 103
10 Hz to 1 MHz in a 1, 3, 10 sequence
and 2 MHz (3 MHz at –6 dB)
Accuracy
1 Hz to 300 kHz
±10%
1 MHz
±25%
2 MHz
+50%, –25%
Selectivity (–60 dB/–3 dB BW ratio)
RBW
≥
300 Hz
<15:1
RBW
≤
100 Hz
<5:1
Video bandwidth
1 Hz to 3 MHz in a 1, 3, 10 sequence
range
Noise sidebands
(see figure 1)
Center Frequency
≤
1 GHz
Offset
Opt. 103
100 Hz
≤
88 dBc/Hz
2
≤
70 dBc/Hz
2
1 kHz
≤
97 dBc/Hz
2
≤
90 dBc/Hz
2
10 kHz
6
≤
113 dBc/Hz
3
≤
113 dBc/Hz
3
30 kHz
6
,
8
≤
113 dBc/Hz
4
≤
113 dBc/Hz
4
100 kHz
7
≤
117 dBc/Hz
5
≤
117 dBc/Hz
5
Residual FM
(zero span, 10 Hz RBW))
<1 Hz pk-pk x N
9
in 20 ms
<0.25 Hz pk-pk x N
9
in 20 ms
(typical)
Option 103
<10 Hz pk-pk x N
9
in 20 ms
Frequency specifications, continued
1. Option 007 extends digitized trace data capability to sweep times <30 ms.
2. Add 5.2 x ((f/1 GHz)–1) for f >1 GHz and f
≤
2.9 GHz
3. Add 2.5 x ((f/1 GHz)–1) for f >1 GHz and f
≤
2.9 GHz
4. Add 3.0 dB x ((f/1 GHz)–1) for f >1 GHz and f
≤
2.9 GHz
5. Add 2 dB for f >1 GHz and f
≤
2.9 GHz
6. RBW
≤
1 k or Span
≤
745 kHz
7. RBW
≥
3 k or Span >745 kHz
8. Not specified at 30 kHz offset for 8564E and 8565E
9. N = harmonic mixing mode number