Sweep time

Range
Span = 0 Hz

50 µs to 6000 s

Span ±100 Hz

RBW

300 Hz

50 ms to 2000 s

RBW

100 Hz

50 ms to 100 ks

Accuracy (Span = 0 Hz)
Sweep time

30 ms

±1% (digitized trace data)

Sweep time <30 ms

±10% (analog trace data)

(non-Option 007)
Sweep time <30 ms

±0.1% (digitized trace data)

(Option 007

1

)

Sweep trigger

delayed, free run, single,
line, video, external

Resolution bandwidth

Range (–3 dB)

1 Hz to 1 MHz in a 1, 3, 10 sequence
and 2 MHz (3 MHz at –6 dB)

Option 103

10 Hz to 1 MHz in a 1, 3, 10 sequence
and 2 MHz (3 MHz at –6 dB)

Accuracy

1 Hz to 300 kHz

±10%

1 MHz

±25%

2 MHz

+50%, –25%

Selectivity (–60 dB/–3 dB BW ratio)
RBW

300 Hz

<15:1

RBW

100 Hz

<5:1

Video bandwidth

1 Hz to 3 MHz in a 1, 3, 10 sequence

range

Noise sidebands

(see figure 1)

Center Frequency

1 GHz

Offset

Opt. 103

100 Hz

88 dBc/Hz

2

70 dBc/Hz

2

1 kHz

97 dBc/Hz

2

90 dBc/Hz

2

10 kHz

6

113 dBc/Hz

3

113 dBc/Hz

3

30 kHz

6

,

8

113 dBc/Hz

4

113 dBc/Hz

4

100 kHz

7

117 dBc/Hz

5

117 dBc/Hz

5

Residual FM

(zero span, 10 Hz RBW))

<1 Hz pk-pk x N

9

in 20 ms

<0.25 Hz pk-pk x N

9

in 20 ms

(typical)

Option 103

<10 Hz pk-pk x N

9

in 20 ms

Frequency specifications, continued

1. Option 007 extends digitized trace data capability to sweep times <30 ms.
2. Add 5.2 x ((f/1 GHz)–1) for f >1 GHz and f

2.9 GHz

3. Add 2.5 x ((f/1 GHz)–1) for f >1 GHz and f

2.9 GHz

4. Add 3.0 dB x ((f/1 GHz)–1) for f >1 GHz and f

2.9 GHz

5. Add 2 dB for f >1 GHz and f

2.9 GHz

6. RBW

1 k or Span

745 kHz

7. RBW

3 k or Span >745 kHz

8. Not specified at 30 kHz offset for 8564E and 8565E
9. N = harmonic mixing mode number