beautypg.com

Frequency specifications, agilent 8560 e-series, Frequency reference, Frequency readout accuracy – Atec Agilent-8564E User Manual

Page 2: Frequency counter accuracy, Frequency span, Frequency range

background image

2

Frequency band

Harmonic mixing mode (N)

30 Hz to 2.9 GHz

1

2.75 GHz to 6.46 GHz

1

5.86 GHz to 13.2 GHz

2

12.4 GHz to 26.8 GHz

4

26.4 GHz to 31.15 GHz

4

31.0 GHz to 50 GHz

8

Frequency reference

Option 103

Temperature stability

3

±1 x 10

-8

±1x10

-6

Aging (per year)

±1 x 10

-7

±2x10

-6

(per day nom.)

±5 x 10

-10

(4)

Initial achievable accuracy

±2.2 x 10

-8

±1 x 10

-6

Short-term warm-up

accuracy factors (nominal)

5 minute

±1 x 10

-7

15 minute

±1 x 10

-8

Frequency readout accuracy

(Start, stop, center, and marker frequency functions)
Span >2 MHz x N

5

±(freq readout x freq ref accuracy

6

+5% x span +15% x RBW +10 Hz)

Span

2 MHz x N

5

±(freq readout x freq ref accuracy

6

+1% x span +15% x RBW +10 Hz)

Frequency counter accuracy

Marker count accuracy

±(marker freq x freq ref

(S/N

25 dB)

accuracy

6

+2 Hz x N

5

+1 LSD of counter)

Accuracy at 1 GHz

±225 Hz (5 minute warm-up)

7

(25 °C, 1 yr aging, marker

±135 Hz (15 minute warm-up)

7

resolution = 1 Hz)

±3003 Hz (Option 103)

Delta count accuracy

±(delta freq x freq ref accuracy

6

(S/N

25 dB)

+4 Hz x N

5

+2 LSD)

Counter resolution

Selectable from 1 Hz to 1 MHz

Frequency span

Range

0, 100 Hz to full span
(100 Hz x N

10

when using

external mixers)

Accuracy
Span >2 MHz x N

5

±5%

Span

2 MHz x N

5

±1%

Frequency specifications, Agilent 8560 E-series

Frequency range

8560E

8561E

8562E

8563E

8564E

8565E

Internal

30 Hz2 to

30 Hz2 to

30 Hz2 to

30 Hz1 to

30 Hz1 to

30 Hz1 to

mixing

2.9 GHz

6.5 GHz

13.2 GHz

26.5 GHz

40 GHz

50 GHz

External

18 GHz to

18 GHz to

18 GHz to

18 GHz to

18 GHz to

18 GHz to

mixing

325 GHz

325 GHz

325 GHz

325 GHz

325 GHz

325 GHz

1. 8563E, 8564E, 8565E require Option 006 for operation below 9 kHz.
2. 8560E, 8561E, 8562E minimum frequency in AC coupled mode is 100 kHz. In DC

coupled mode minimum frequency is 30 Hz.

3. -10 °C to +55 °C, referenced to 25 °C
4. After 7-day warm-up
5. N = harmonic mixing mode number
6. Frequency reference accuracy = aging x time since last adjustment + initial

achievable accuracy + temperature stability

7. Short term warm-up accuracy factors have been included in this calculation.