Frequency specifications, agilent 8560 e-series, Frequency reference, Frequency readout accuracy – Atec Agilent-8564E User Manual
Page 2: Frequency counter accuracy, Frequency span, Frequency range
![background image](/manuals/318705/2/background.png)
2
Frequency band
Harmonic mixing mode (N)
30 Hz to 2.9 GHz
1
2.75 GHz to 6.46 GHz
1
5.86 GHz to 13.2 GHz
2
12.4 GHz to 26.8 GHz
4
26.4 GHz to 31.15 GHz
4
31.0 GHz to 50 GHz
8
Frequency reference
Option 103
Temperature stability
3
±1 x 10
-8
±1x10
-6
Aging (per year)
±1 x 10
-7
±2x10
-6
(per day nom.)
±5 x 10
-10
(4)
Initial achievable accuracy
±2.2 x 10
-8
±1 x 10
-6
Short-term warm-up
accuracy factors (nominal)
5 minute
±1 x 10
-7
15 minute
±1 x 10
-8
Frequency readout accuracy
(Start, stop, center, and marker frequency functions)
Span >2 MHz x N
5
±(freq readout x freq ref accuracy
6
+5% x span +15% x RBW +10 Hz)
Span
≤
2 MHz x N
5
±(freq readout x freq ref accuracy
6
+1% x span +15% x RBW +10 Hz)
Frequency counter accuracy
Marker count accuracy
±(marker freq x freq ref
(S/N
≥
25 dB)
accuracy
6
+2 Hz x N
5
+1 LSD of counter)
Accuracy at 1 GHz
±225 Hz (5 minute warm-up)
7
(25 °C, 1 yr aging, marker
±135 Hz (15 minute warm-up)
7
resolution = 1 Hz)
±3003 Hz (Option 103)
Delta count accuracy
±(delta freq x freq ref accuracy
6
(S/N
≥
25 dB)
+4 Hz x N
5
+2 LSD)
Counter resolution
Selectable from 1 Hz to 1 MHz
Frequency span
Range
0, 100 Hz to full span
(100 Hz x N
10
when using
external mixers)
Accuracy
Span >2 MHz x N
5
±5%
Span
≤
2 MHz x N
5
±1%
Frequency specifications, Agilent 8560 E-series
Frequency range
8560E
8561E
8562E
8563E
8564E
8565E
Internal
30 Hz2 to
30 Hz2 to
30 Hz2 to
30 Hz1 to
30 Hz1 to
30 Hz1 to
mixing
2.9 GHz
6.5 GHz
13.2 GHz
26.5 GHz
40 GHz
50 GHz
External
18 GHz to
18 GHz to
18 GHz to
18 GHz to
18 GHz to
18 GHz to
mixing
325 GHz
325 GHz
325 GHz
325 GHz
325 GHz
325 GHz
1. 8563E, 8564E, 8565E require Option 006 for operation below 9 kHz.
2. 8560E, 8561E, 8562E minimum frequency in AC coupled mode is 100 kHz. In DC
coupled mode minimum frequency is 30 Hz.
3. -10 °C to +55 °C, referenced to 25 °C
4. After 7-day warm-up
5. N = harmonic mixing mode number
6. Frequency reference accuracy = aging x time since last adjustment + initial
achievable accuracy + temperature stability
7. Short term warm-up accuracy factors have been included in this calculation.