Specifications, Continued), Ask/fsk modulation analysis (option dma) – Atec Agilent-N9342C User Manual
Page 8
8
Option specifications (continued)
Supplemental information
AM/FM modulation analysis (Option AMA)
Frequency range
10 MHz to 7 GHz
Carrier power accuracy
± 1.8 dB
Nominal
Carrier power range
–30 to +10 dBm
100 kHz to 2 MHz
–30 to +20 dBm
2 MHz to 7 GHz
Carrier power displayed resolution
0.01 dBm
AM measurement
Modulation rate
20 Hz to 100 kHz
Accuracy
1 Hz
Nominal (modulation rate < 1 kHz)
< 0.1% modulation rate
Nominal (modulation rate > 1 kHz)
Depth
5 to 95%
Accuracy
± 4%
Nominal
FM measurement
Modulation rate
20 Hz to 200 kHz
Accuracy
1 Hz
Nominal (modulation rate < 1 kHz)
< 0.1% modulation rate
Nominal (modulation rate > 1 kHz)
Depth
20 Hz to 400 kHz
Accuracy
± 4%
Nominal
ASK/FSK modulation analysis (Option DMA)
Frequency range
2.5 MHz to 6 GHz
Carrier power accuracy
± 2 dB
Nominal
Carrier power range
–30 to +20 dBm
Nominal
Carrier power displayed resolution
0.01 dBm
ASK measurement
Symbol rate range
100 Hz to 100 kHz
Modulation depth/index
5 to 95%
Accuracy
± 4%
Nominal
Displayed resolution
0.1%
FSK measurement
FSK deviation
100 Hz to 400 kHz
Symbol rate range
100 Hz to 20 kHz
1 ≤ ß* ≤ 20
20 to 50 kHz
1 ≤ ß ≤ 8
50 to 100 kHz
1 ≤ ß ≤ 4
Accuracy
± 4%
Nominal
Displayed resolution
0.01 Hz
* ß is the ratio of frequency deviation to symbol rate (deviation/rate)
Specifications
(continued)