Specifications, Continued) – Atec Agilent-N9342C User Manual
Page 11
11
Option specifications (continued)
Supplemental
information
Baseband input (Option BB1)
Frequency range
9 kHz to 12 MHz, usable to 3 kHz
Frequency span
100 Hz to 11.997 MHz
Frequency resolution
1 Hz
Measurement range
9 kHz to 2 MHz
Displayed average noise level
(DANL) to +10 dBm
2 MHz to 12 MHz
Displayed average noise level
(DANL) to +20 dBm
Overall amplitude accuracy 9 kHz to 100 kHz
± 2.5 dB
20 to 30 °C, 30 to 70% RH, peak
detector, preamp off, input signal
–50 to 0 dBm, 95% percentile
100 kHz to 12 MHz
± 1.5 dB
Normalized to 1 Hz
Minimum RBW
Displayed average noise
level
9 kHz to 100 kHz, nominal
–145 dBm
9 kHz to 100 kHz, nominal
–135 dBm
RMS detector, trace averaging
> 40, 0 dB input attenuation,
input terminated 50 Ω, reference
level < –35 dBm
100 kHz to 12 MHz, –155 dBm,
typical –158 dBm
100 kHz to 12 MHz, –145 dBm,
typical –148 dBm
Phase noise
10 kHz nominal –118 dBc/Hz
Center frequency 10 MHz
30 kHz nominal –120 dBc/Hz
100 kHz nominal –127 dBc/Hz
> 200 kHz nominal –130 dBc/Hz
Residual responses
< –120 dBm nominal
9 kHz to 12 MHz
Second harmonic distortion < –55 dBc nominal
Input frequency > 100 kHz, mixer
signal level at –30 dBm, input
attenuation 0 dB, preamp off,
reference level –30 dBm,
20 to 30 °C
Third order intermodulation
distortion
< –55 dBc nominal
Two –20 dBm tones at input
mixer, spaced by 100 kHz, input
attenuation 0 dB, preamp off,
reference level ≥ –20 dBm,
20 to 30 °C
VSWR
< 2.0:0 nominal
10 or 20 dB attenuation, 300 kHz
to 12 MHz
Specifications
(continued)