beautypg.com

Specifications, Continued) – Atec Agilent-N9342C User Manual

Page 6

background image

6

Specifications

(continued)

Sweep specifications

Supplemental information

Sweep time

Range

2 ms to 1000 s

Span ≥ 100 Hz

600 ns to 200 s

Span = 0 Hz (zero span)

Sweep mode

Continuous, single

Sweep time rule

Accuracy, speed

Trigger source

Free run, video, external, RF burst

Trigger slope

Selectable positive or negative edge

Trigger delay

± 12 ms to ± 12 s

Nominal, span = 0 Hz (zero span)

Front panel input/output

Supplemental information

RF input

Connector and impedance

Type-N female, 50 Ω

Nominal

10 MHz reference/external trigger input

Reference input frequency

10 MHz

Reference input amplitude

0 to +10 dBm

Trigger voltage

5 V TTL level

Nominal

Connector

BNC female, 50 Ω

Nominal

Probe power

Voltage/current

+15 Vdc, ± 7% at 0 to 150 mA (nominal)
–12.6 Vdc, ± 10% at 0 to 150 mA (nominal)
GND

Connectivity

USB host

USB Type-A female, compatible with USB

2.0 full speed

USB device

USB Type-mini AB female, compatible with

USB 2.0 full speed

LAN

RJ-45, 10 Base-T

General specifications

Supplemental information

Display

Resolution

640 pixels x 480 pixels

Size and type

170 mm (6.5 in) TFT color display

Internal memory

System memory

64 MB

For system use. Not user accessible.

User memory

64 MB

User accessible. Able to store about 14,000

traces.

Languages

On-screen GUI

English, Simplified Chinese, Traditional

Chinese, French, German, Italian,

Japanese, Korean, Russian, Spanish,

Portuguese