Specifications, Continued), Maximum safe input level – Atec Agilent-N9342C User Manual
Page 4: Displayed average noise level, Level display range
![background image](/manuals/318691/4/background.png)
4
Specifications
(continued)
Amplitude specifications
Supplemental
information
Measurement range
100 kHz to 2 MHz
Displayed average noise level (DANL) to +10 dBm
Preamp off
2 MHz to 7 GHz
Displayed average noise level (DANL) to +20 dBm
Input attenuator range
0 to 50 dB, in 1 dB steps
Maximum safe input level
Average continuous
power
+33 dBm, 3 minutes maximum
Input attenuator setting
≥ 20 dB, 2 MHz to 7 GHz
DC voltage
± 50 VDC maximum
Displayed average noise level
1
Preamp off
Normalized to 1 Hz
Minimum RBW
100 kHz to 1 MHz
–108 dBm, typical –127 dBm
–98 dBm, typical –117 dBm
Reference level
≤ –50 dBm
1 to 10 MHz
–128 dBm, typical –146 dBm
–118 dBm, typical –136 dBm
10 to 500 MHz
–142 dBm, typical –146 dBm
–132 dBm, typical –136 dBm
500 MHz to 2.5 GHz
–141 dBm, typical –145 dBm
–131 dBm, typical –135 dBm
2.5 to 4 GHz
–140 dBm, typical –144 dBm
–130 dBm, typical –134 dBm
4 to 6 GHz
–138 dBm, typical –142 dBm
–128 dBm, typical –132 dBm
6 to 7 GHz
–136 dBm, typical –140 dBm
–126 dBm, typical –130 dBm
Preamp on
100 kHz to 1 MHz
–131 dBm, typical –150 dBm
–121 dBm, typical –140 dBm
Reference level
≤ –70 dBm
1 to 10 MHz
–148 dBm, typical –163 dBm
–138 dBm, typical –153 dBm
10 to 500 MHz
–161 dBm, typical –164 dBm
–151 dBm, typical –154 dBm
500 MHz to 2.5 GHz
–159 dBm, typical –162 dBm
–149 dBm, typical –152 dBm
2.5 to 4 GHz
–158 dBm, typical –161 dBm
–148 dBm, typical –151 dBm
4 to 6 GHz
–155 dBm, typical –158 dBm
–145 dBm, typical –148 dBm
6 to 7 GHz
–150 dBm, typical –154 dBm
–140 dBm, typical –144 dBm
Level display range
Log scale
10 to 100 dB, 10 divisions displayed, 1, 2, 5, 10 dB/division
Linear scale
0 to 100%, 10 divisions displayed
Scale units
dBm, dBmV, dBμV, W, V, dBmV EMF, dBμV EMF, V EMF
Sweep (trace) points
461
Marker level readout
resolution
Log scale
0.01 dB
Linear scale ≤ 1% of signal level
Nominal
Detectors
Normal, positive peak, sample, negative peak, average (video,
RMS, voltage)
Number of traces
4
1. RMS detector, trace averaging > 40, 0 dB input attenuation, input terminated 50 Ω, 1 kHz resolution bandwidth, 20 to 30 °C.