Level accuracy with step attenuator

1

(dB)

Frequency

> +10 dBm

+10 to –10 dBm

–10 to –70 dBm

–70 to –90 dBm

–90 to –110 dBm

250 kHz to 2 GHz

±0.6

±0.6 ±0.7 ±0.8 ±1.4

> 2 to 20 GHz

±0.8

±0.8

±0.9

±1.0

±1.7

> 20 to 40 GHz

±1.0

±0.9

±1.0

±2.0

Resolution

0.01 dB

Temperature stability

0.01 dB/°C (typical)

User flatness correction
Number of points

2 to 1601 points/table

Number of tables

Up to 10,000, memory limited

Path loss

Arbitrary, within attenuator range

Entry modes

Remote power meter

2

, remote bus, manual

(user edit/view)
Output impedance

50

Ω (nominal)

SWR (internally leveled) (typical)
250 kHz to 2 GHz

< 1.4:1

> 2 GHz to 20 GHz

< 1.6:1

> 20 GHz to 40 GHz

< 1.8:1

Leveling modes

Internal leveling, external detector leveling, millimeter
source module, ALC Off

External detector leveling
Range

–0.2 mV to –0.5 V (nominal) (–36 dBm to +4 dBm using

Agilent 33330D/E detector)

Bandwidth

10 kHz (typical) (Note: not intended for pulsed operation)

Maximum reverse power

1/2 Watt (nominal)

1.

Specifications apply in CW and List/Step sweep modes over the 15 to 35º C temperature range, with attenuator
hold off (normal operating mode). Degradation outside this range, for ALC power levels > –10 dBm, is typically
< 0.3 dB. In Ramp sweep mode (with Option 007), specifications are typical. For instruments with type-N
connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz. Level accuracy is not
specified below –110 dBm.

2.

Compatible with Agilent Technologies EPM Series (E4418B and E4419B) power meters.

Measured level accuracy +5 dBm Option 1E1

Frequency (MHz)

1

0.5

0

-0.5

-1

0

5000 10000

15000

20000

Measured level accuracy +5 dBm Option 1E1

Frequency (MHz)

1

0.5

0

-0.5

-1

0

10000 20000

30000 40000

20 GHz level accuracy

40 GHz level accuracy