Output – Atec Agilent-E8247C_57C User Manual
Page 5
5
Power
1
(dBm)
Frequency range
Standard
Option 1EA
20 GHz models
250 kHz to 3.2 GHz
–20 to +13
–20 to +16
250 kHz to 3.2 GHz (with Option 1E6) –20 to +13
–20 to +13
> 3.2 to 20 GHz
–20 to +13
–20 to +20
40 GHz models
250 kHz to 3.2 GHz
–20 to +9
–20 to +15
250 kHz to 3.2 GHz (with Option 1E6) –20 to +9
–20 to +12
> 3.2 to 20 GHz
–20 to +9
–20 to +18
> 20 to 40 GHz
–20 to +9
–20 to +14
20 GHz models with step attenuator (Option 1E1)
250 kHz to 3.2 GHz
–135 to +11
–135 to +15
250 kHz to 3.2 GHz (with Option 1E6) –135 to +11
–135 to +12
> 3.2 to 20 GHz
–135 to +11
–135 to +18
40GHz models with step attenuator (Option 1E1)
250 kHz to 3.2 GHz
–135 to +7
–135 to +14
250 kHz to 3.2 GHz (with Option 1E6) –135 to +7
–135 to +11
> 3.2 to 20 GHz
–135 to +7
–135 to +16
> 20 to 40 GHz
–135 to +7
–135 to +12
Step attenuator
0 dB and 5 to 115 dB in 10 dB steps
3
(Option 1E1)
Attenuator hold range
(Same as max power sweep range)
Minimum
From –20 dBm to maximum specified output power with
step attenuator in 0 dB position. Can be offset using
Option 1E1 attenuator.
Amplitude switching speed
2
CW or analog modulation
< 5 ms (typical)
When using power search
< 25 ms (typical)
Level accuracy
3
(dB)
Frequency
> +10 dBm
+10 to –10 dBm
–10 to –20 dBm
250 kHz to 2 GHz
±0.6
±0.6
±1.4
2 GHz to 20 GHz
±0.8
±0.8
±1.2
> 20 to 40 GHz
±1.0
±0.9
±1.3
Output
1.
Maximum power specification is warranted from 15 to 35º C, and is typical from 0 to 15º C. Maximum
power over the 35 to 55º C range typically degrades less than 2 dB.
2.
To within 0.1 dB of final amplitude within one attenuator range
3.
Specifications apply in CW and List/Step sweep modes over the 15 to 35º C temperature range.
Degradation outside this range, for power levels > –10 dBm, is typically < 0.3 dB. In Ramp sweep mode
(with Option 007), specifications are typical. For instruments with Type-N connectors (Option 1ED), specifications
are degraded typically 0.2 dB above 18 GHz.
20 GHz Models with Option 1EA
26
24
22
20
18
16
0 5000 10000 15000 20000
Frequency (MHz)
40 GHz Models with Option 1EA
26
24
22
20
18
16
0 10000 20000 30000 40000
Frequency (MHz)
20 GHz models with Option 1EA
40 GHz models with Option 1EA
Measured maximum available power
Measured maximum available power