Remote programming general specifications – Atec Agilent-E8247C_57C User Manual
Page 13
13
Interfaces
GPIB (IEEE-488.2,1987) with listen and talk, RS-232,
and 10BaseT LAN interface.
Control languages
SCPI version 1997.0. Also will emulate most applicable
Agilent 836xxB, Agilent 837xxB, and Agilent 8340/41B
commands, providing general compatibility with ATE
systems which include these signal generators.
IEEE-488 functions
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1,
DT0, C0, E2.
ISO compliant
This family of signal generators is manufactured in an
ISO-9001 registered facility in concurrence with Agilent
Technologies commitment to quality.
Power requirements
90 to 132 VAC 50 to 60 Hz or 365 to 435 Hz;
195 to 267 VAC 50 to 60 Hz, (automatically selected),
300 W maximum.
Operating temperature range
0 to 55 °C
Storage temperature range
1
–40 to 71 °C
Shock and vibration
Operating random vibration
5 to 500 Hz, 0.21 g rms
Survival swept sine vibration
5 to 500 Hz, 0.75 g
Survival random vibration
5 to 500 Hz, 2.09 g rms
Functional shock (half-sine, 30 g, 11 ms) Meets the requirements of MIL-PRF-28800F for class
and bench drop test
3 equipment.
EMC
Meets the conducted and radiated interference
and immunity requirements of IEC/EN 61326-1.
Meets radiated emission requirements of
CISPR Pub 11/1997 Group 1 class A.
Storage registers
Memory is shared by instrument states, user data files,
sweep list files, and waveform sequences. Depending
on the number and size of these files, up to 800 storage
registers and 10 register sequences are available.
Security
Display blanking.
Compatibility
Agilent Technologies 83550 Series millimeter heads,
Agilent Technologies 8757D Scalar Network Analyzers,
Agilent Technologies EPM Series Power Meters.
Self-test
Internal diagnostic routine tests most modules
(including microcircuits) in a preset condition. For each
module, if its node voltages are within acceptable limits,
then the module “passes” the test.
Weight
< 22 kg (48 lb.) net, < 30 kg (68 lb.) shipping.
Dimensions
178 mm H x 426 mm W x 498 mm D
(7” H x 16.8” W x 19.6” D in.).
Recommended calibration cycle
24 months
Remote programming
General specifications
1.
Storage below –20 °C instrument states may be lost.