Digital sweep

Ramp (analog) sweep
(Option 007)

1

Operating modes

Step sweep of frequency or amplitude or both (start to stop)
List sweep of frequency or amplitude or both (arbitrary list)

Sweep range
Frequency sweep

Within instrument frequency range

Amplitude sweep

Within attenuator hold range

Dwell time

1 ms to 60 s

Number of points

2 to 1601 (step sweep)
2 to 1601 per table (list sweep)

Triggering

Auto, external, single, or GPIB

Operating modes

Synthesized frequency sweep
(start/stop), (center/span), (swept CW)
Power (amplitude) sweep (start/stop)
Manual sweep
RPG control between start and stop frequencies
Alternate sweep
Alternates successive sweeps between current and stored states

Sweep span range

Settable from minimum

2

to full range

Maximum sweep rate

Start frequency

Maximum sweep rate Max span for

100ms sweep

250 kHz to < 0.5 GHz

25 MHz/ms

2.5 GHz

0.5 to < 1 GHz

50 MHz/ms

5 GHz

1 to < 2 GHz

100 MHz/ms

10 GHz

2 to < 3.2 GHz

200 MHz/ms

20 GHz

≥ 3.2 GHz

400 MHz/ms

36.8 GHz

Frequency accuracy

± 0.05% of span ± timebase (at 100 ms sweep time, for
sweep spans less than maximum values given above)
Accuracy improves proportionally as sweep time increases

3

Sweep time

(forward sweep, not including bandswitch and retrace intervals)

Resolution

1 ms

Manual mode

Settable 10 ms to 99 seconds

Auto mode

Set to minimum value determined by maximum sweep rate
and 8757D setting

Triggering

Auto, external, single, or GPIB

Markers

10 independent continuously variable frequency markers

Display

Z-axis intensity or RF amplitude pulse

Functions

M1 to center, M1/M2 to start/stop, marker delta

Two-tone (master/slave)
measurements

4

Two PSG’s can synchronously track each other, with
independent control of start/stop frequencies

Network analyzer compatibility

Fully compatible with Agilent 8757D scalar network analyzer

5

Also useable with Agilent 8757A/C/E scalar network analyzers
for making basic swept measurements.

6

1.

During Ramp sweep operation, AM and Pulse Modulation are useable but not specified; FM, Phase
Modulation, Wideband AM and I/Q modulation are not useable.

2.

Minimum settable sweep span is proportional to carrier frequency and sweep time. Actual sweep span
may be slightly different than desired setting for spans less than [0.00004% of carrier frequency or
140 Hz] x [sweep time in seconds]. Actual span will always be displayed correctly.

3.

Typical accuracy for sweep times > 100 ms can be calculated from the equation:
[(0.005% of span)/(sweep time in seconds)] ± timebase. Accuracy is not specified for sweep times < 10 ms.

4.

For Master/Slave operation use Agilent Technologies part #8120-8806 Master/Slave interface cable.

5.

When measuring low-pass devices in AC mode, dynamic range may be reduced up to 10dB below 3.2 GHz

6.

GPIB system interface is not supported with 8757A/C/E, only with 8757D. As a result, some features of
8757A/C/E, such as frequency display, pass-through mode, and alternate sweep, do not function with
PSG signal generators.