General characteristics – Atec Agilent-N6030A User Manual
Page 10
10
Power
Supply
Typical operation (watts)
+3.3 VDC
11.2
+5 VDC
22
+12 VDC
5
-12 VDC
5
Total Power
43.2
Environmental
Samples of this product have been
type tested in accordance with the
Agilent Environmental Test Manual
and verified to be robust against the
environmental stresses of Storage,
Transportation, and End-use; those
stresses include but are not limited to
temperature, humidity, shock vibra-
tion, altitude, and power line condi-
tions. Test methods are aligned with
IEC 60068-2 and levels are similar to
MIL-PRF-28800F Class 3.
Operating temperature:
0 to +55 degrees C
Storage temperature:
-20 to +70 degrees C
Relative Humidity:
Type tested:
10 to 90% at 40 degrees C
(non-condensing)
Altitude:
0 to 2000m (6500 ft) above sea level
Safety
Designed for compliance to IEC
61010-1:2001
EMC
Meets the conducted and radiated
emissions and immunity requirements
of IEC 61326:2002 when tested with
EMC shielded filler panels (Agilent
P/N N6030-80007, kit of 6) separating
the controller and the N6030A
module, and in all open slots. The RFI
gaskets must be oriented to the right.
Weight
1.14 kg (2.5 lb)
Security
All user data stored in volatile memory
Dimensions
3U, 4 slot CompactPCI module
8.1 x 13 x 21.6 cm
(3.2 x 5.1 x 8.5 inches)
ISO compliance
This modular instrument is
manufactured in an ISO-900
registered facility in concurrence with
Agilent Technologies, Inc.
commitment to quality.