Power Sensor Specifications

Frequency Range

CW Dynamic Range

(dBm)

SWR

Rise Time

1

(ms)

Sensor Linearity

6

RF

Connector

2

Standard Diode Sensors

MA2472D

10 MHz to18 GHz

–70 to +20 CW Mode
–30 to +20 (with ML2480A)

–34 to +20 (with ML2490A)

Pulse/Modulated Mode

<1.17; 10 to 150 MHz (MA2472D only)

<1.90; 10 to 50 MHz

<1.17; 50 to 150 MHz

<1.12; 0.15 to 2 GHz

<1.22; 2 to 12.4 GHz

<1.25; 12.4 to 18 GHz

<1.35; 18 to 32 GHz

<1.50; 32 to 40 GHz

<1.63; 40 to 50 GHz

<0.004

1.8%, <18 GHz

2.5%, 40 GHz

3.5%, <50 GHz

For MA2475D

4

N (m)

MA2473D

10 MHz to 32 GHz

K (m)

MA2474D

10 MHz to 40 GHz

K (m)

MA2475D

10 MHz to 50 GHz

V (m)

Temperature accuracy: <1% < 40 GHz, <1.5% <50 GHz, 5 to 50º C

High Accuracy Diode Sensors

MA2442D

MA2444D

10 MHz to 18 GHz

10 MHz to 40 GHz

–67 to +20

<1.90; 10 MHz to 50 MHz

<1.17; 10 MHz to 150 MHz (MA2442D only)

<1.17; 50 MHz to 150 MHz

<1.08; 150 MHz to 2 GHz

<1.16; 2 GHz to 12.4 GHz

<1.21; 12.4 GHz to 18 GHz

<1.29; 18 GHz to 32 GHz

<1.44; 32 GHz to 40 GHz

<1.50; 40 GHz to 50 GHz

<0.004

1.8%, <18 GHz

2.5%, <40 GHz

3.5%, <50 GHz

For MA2445D

5

N (m)

K (m)

MA2445D

10 MHz to 50 GHz

V (m)

Temperature accuracy: <1% < 40 GHz, 5 to 50ºC

Universal Power Sensors

MA2481D

MA2482D

10 MHz to 6 GHz

10 MHz to 18 GHz

–60 to +20

<1.17; 10 to 150 MHz

<1.12; 0.15 to 2 GHz

<1.22; 2 to 12.4 GHz

<1.25; 12.4 to 18 GHz

<0.004

(with option 1 only)

10 MHz to 6 GHz

3% –60 to +20 dBm
6 to 18 GHz

3% –60 to 0 dBm

3.5% 0 to +20 dBm

(1.8% CW with option 1)

N (m)

Temperature accuracy: <1%, 5 to 50º C
MA2480/01

Adds fast CW mode to Universal Power Sensors for high speed measurements of CW signal plus TDMA and pulse measurements

Wideband Sensors

MA2490A

3

50 MHz to 8 GHz

–60 to +20 CW Mode
–25 to +20 (with ML2480A)

–30 to +20 (with ML2490A)

Pulse/Modulated Mode

<1.17; 50 to 150 MHz

<1.12; 0.15 to 2.5 GHz

<1.22; 2.5 to 8 GHz

<18 ns

<7% 50 to 300 MHz

<3.5% 0.3 to 8 GHz

N (m)

MA2491A

3

50 MHz to 18 GHz

<1.17; 50 to 150 MHz

<1.12; 0.15 to 2.5 GHz

<1.22; 2.5 to 12.4 GHz

<1.25; 12.4 to 18 GHz

<7% 50 to 300 MHz

<3.5% 0.3 to 18 GHz

N (m)

Temperature accuracy: <1% 10 to 45º C

Pulse Sensor

MA2411B

Requires 1 GHz calibrator

option 15 to be fitted

300 MHz to 40 GHz –20 to +20 dBm

<1.15; 0.3 to 2.5 GHz

<1.35; 2.5 to 26 GHz

<1.50; 26 to 40 GHz

<8 ns

<18 ns when used

with ML2487/8A

<4.5% 0.3 to 18 GHz

<7% 18 to 40 GHz

K (m)

Temperature accuracy: <2% 10 to 45º C

1

0.0 dBm, room temperature with standard 1.5m sensor cable.

2

Each MA2400A/D Series sensor incorporates precision RF connectors with hexagon

coupling nut for attachment by industry standard torque wrench.

3

MA2490/1A and MA2411B sensors must be used with ML2480A or

ML24950A Power Meters.

4

For Linearity on MA2475D only applicable to –70 to +15 dBm.

5

For Linearity on MA2445D only applicable to -67 to +15 dBm.

6

Sensor linearity specifications are ± value.

Pulse/modulated performance only specified with 1.5m sensor cable length option

ML2400A-20 supplied as standard with the power meter.