beautypg.com

Power meter specifications – Atec Anritsu-MA2445A User Manual

Page 3

background image

3

Power Meter Specifications

ML2430A Series

ML2480A Series

ML2490A Series

Comments

ML2437A

ML2438A

ML2487A

ML2488A

ML2495A

ML2496A

Signal Inputs

1

2

1

2

1

2

Frequency range

100 kHz to 65 GHz (sensor dependent)

Dynamic range

–70 to +20 dBm (dependent on sensor, external coupler or attenuator)

Continuous or Peak

Performance

100 kHz (Profile mode)

Pulse/Modulated mode

20 MHz with MA2491A sensor
CW mode

17 kHz ranges 1–4

35 Hz range 5

Pulse/Modulated mode

>65 MHz range 7

>38 MHz range 8

>16 MHz range 9

(Repetitive Sampling)
20 MHz (One shot)
Combined B/W (with MA2411B sensor)

>39 MHz range 7

>29 MHz range 8

>12 MHz range 9

MA2411B nominal Bandwidth = 50 MHz
CW mode

17 kHz range 1,2,3,4

36 Hz range 5

Nominal Video BW

31.25 kS/s

Auto/Manual
CW Mode

75 kS/s
Pulse/Modulated Mode

31.25 kS/s to 64 MS/s

(dependent on trigger capture time)

Auto/Manual
CW Mode

75 kS/s
Pulse/Modulated Mode

31.25 kS/s to 62.5 MS/s Continuous Sampling

(Trigger capture time 3.2 µs to 7s,

200 data points)
1 GS/s Random Repetitive Sampling

(Trigger capture time 50 ns - 3.2 ns,

200 data points)
Conflicts between selected settings and other

instrument settings are indicated through user

warnings (displayed and GPIB)

Sampling rate

N/A

<18 ns

(with MA2411B sensor)

Typical 8 ns, Maximum 12 ns

(with MA2411B sensor)

Fall-time typically 11 ns

System rise-time

(10% to 90% at +10 dBm)

N/A

10% to 90% Rise-time measurement of –20 dBm to +20 dBm Peak power (with MA2491A)

Rise-time measurement

dynamic range

N/A

≤3% in linear power at +10 dBm

Overshoot

(Pulse/Modulated mode)

N/A

±100 ppm (1 GHz clock for RRRS mode)

Sampling Rate

Clock Accuracy

Accuracy
(Defined by

uncertainty

calculations with

relevant sensor

and source match

conditions)

<0.5%

CW Mode

<0.5% (±0.02 dB absolute Accuracy, ±0.04 dB relative Accuracy)
Pulse/Modulated Mode

<0.8% Nominal range 7, 8

Instrumentation Accuracy

<0.5% of full scale in most sensitive range,

measured over one hour with maximum

averaging after one hour warm up at constant

temperature

Equivalent Noise Power

(256 Moving Average)

MA2472D MA2491A

Range 1

0.5 µW

2 µW

Range 2

50 nW

100 nW

Range 3

0.5 nW

2 nW

Range 4

0.2 nW

1 nW

Range 5

50 pW

0.5 nW

(CW mode)
Range 7

20 µW

68 µW

Range 8

3 µW

10 µW

Range 9

2 µW

8 µW

(Pulse/Modulated mode)

Equivalent Noise Power

(256 Moving Average)

MA2472D MA2491A

Range 1

0.5 µW

2 µW

Range 2

50 nW

100 nW

Range 3

0.5 nW

2 nW

Range 4

0.2 nW

1 nW

Range 5

50 pW

0.5 nW

(CW mode)
Range 7

5 µW

15 µW

Range 8

1 µW

5 µW

Range 9

0.5 µW

2 µW

(Pulse/Modulated mode)

Zero set and Drift