Atec Agilent-60504B User Manual
Page 6
![background image](/manuals/318499/6/background.png)
7
Table 60504-2. Programming Ranges (continued)
Function
Front Panel
Front Panel
HPSL Command
Range of Values
Key
Display
(Short Form)
Constant Resistance
Set Slew Rate
(shift)
Low Range
V:SLW value
"VOLT:SLEW value"
same as voltage slew
Middle/High Range
C:SLW value
"CURR:SLEW value"
same as current slew
Set Transient Level
R:TLV value
"RES:TLEV value"
same as main level
*Set Triggered Level
"RES:TRIG value"
same as main level
Constant Voltage
Set Main Level
VOLT value
"VOLT value"
0 to 60 V
Set Slew Rate
(shift)
V:SLW value
"VOLT:SLEW value"
0.001 to 0.5 (V/
µ
s)
Set Transient Level
V:TLV value
"VOLT:TLEV value"
same as main level
*Set Triggered Level
"VOLT:TRIG value"
same as main level
Transient Operation
Set Frequency
FREQ value
"TRAN:FREQ value"
0.25 Hz to 10 kHz
Set Duty Cycle
(shift)
DCYCLE value
"TRAN:DCYC value"
3-97% (0.25 Hz-1 kHz)
6-94% (1 kHz-10 kHz)
*Set Pulse Width
"TRAN:TWID value"
0.00005 to 4 s
Trigger Operation
*Set Trigger Period
"TRIG:TIM value"
0.000008 to 4 s
Current Protection
*Set Current Level
"CURR:PROT value"
0 to 122.4 A
*Set Delay Time
"CURR:PROT:DEL value"
0 to 60 s
*Can only be programmed remotely via the GPIB.
Table 60504-3. Factory Default Settings
Function
Settings
Function
Setting
CURR level
0 A
Mode (CC, CR, CV)
CC
CURR transient level
0 A
Input (on/off)
on
*CURR slew rate
2 A/
µ
s
Short (on/off)
off
CURR range
120 A
Transient operation (on/off)
off
*CURR protection (on/off)
off
***TRAN mode
continuous
**CURR protection level
122.4 A
(continuous, pulse, toggle)
**CURR protection delay
15 s
TRAN frequency
1 kHz
TRAN duty cycle
50%
RES level
500
Ω
**TRAN pulse width
0.5 ms
RES transient level
500
Ω
RES range
500
Ω
**TRIG source
hold
(bus, external, hold, timer, line)
VOLT level
60 V
**TRIG period
0.001 s
VOLT transient level
60 V
**PORT0 output (on/off)
off (logic 0)
VOLT slew rate
5 V/
µ
s
**CAL mode (on/off)
off
The *RST command resets the CURR slew rate to 0.83 A/
µ
, not to the factory default.
**Can only be programmed remotely via the GPIB.
***Continuous transient mode is the only mode available at the front panel. Pulsed, toggled, and continuous modes can
all be programmed remotely via the GPIB.