6

Range

Transient Current Level

Current Slew Rate

Overshoot*

12 A

6 A

0.5 A/

µ

s, 1 A/

µ

s

5%

6 A

0.2 A/

µ

s to 0.2 A/

µ

s

0

12 A

1 A/

µ

s

2%

12 A

0.2 A/ms to 0.5 A/

µ

s

0

*Overshoot may be higher during the first five seconds of programming if unit has been operating at full
current. Overshoot values assume a total inductance of l

µ

H, or less, in the load leads connected to the

D.U.T.

Source Turn-On Current Overshoot:

Less than 10% of final value (in CC and CR modes when connected to power

supplies with voltage rise times of greater than 500

µ

s).

Programmable Short Circuit:

0.17

(0.012

typical)

Programmable Open Circuit:

20 k

(typical)

Drift Stability

(over an 8 hour interval):

Current:

±0.03% ± 20 mA

Voltage:

±0.01% ± 10 mV

Reverse Current Capacity:

120 A when unit is on; 60 A when unit is off

Weight:

5.4 kg (12 lbs.)

Table 60504-2. Programming Ranges

Function

Front Panel

Front Panel

HPSL Command

Range of Values

Key

Display

(Short Form)

Constant Current

Set Range

C:RNG value

"CURR:RANG value"

Low Range

0 and

12 A

High Range

> l2 A and

120 A

Set Main Level

CURR value

"CURR value"

Low Range

0 to 12 A

High Range

0 to 120 A

Set Slew Rate

(shift)

C:SLW value

"CURR:SLEW value"

Low Range

0.0002 to 1 (A/

µ

s)

High Range

0.002 to 10 (A/

µ

s)

Set Transient Level

C:TLV value

"CURR:TLEV value"

same as main level

*Set Triggered Level

"CURR:TRIG value"

same as main level

Constant Resistance

Set Range

R:RNG value

"RES:RANG value"

Low Range

0 and

0.5

Middle Range

> 0.5

and

500 k

High Range

>500

and

5 k

Set Main Level

RES value

"RES value"

Low Range

0 to 0.5

Middle Range

0.5

to 500

High Range

5

to 5 k