Table 60501-4. Calibration Information

Ranges and
Calibration Points

Variables

Variables

Value

Power Supply

Settings

Current

Shunt

High Current Range

Hi_curr_rng

30

5 V/31 A

100 A

High Current Offset

Hi_curr_offset

0.013

Low Current Range

Lo_curr_rng

3

5 V/10 A

15 A

Low Current Offset

Lo_curr_offset

0.013

Voltage Range

N/A

N/A

61 V/2 A

N/A

Voltage Hi point

Volt_hipt

60

Voltage Lo point

Volt_lopt

2.7

Low Resistance Range

Lo_res_rng

2

15 V/5.5 A

15 A

Low Resistance Hi point

Lo_res_hipt

1.9

Low Resistance Lo point

Lo_res_lopt

0.067

Middle Resistance Range

Mid_res_rng

20

10.9 V/8 A

15 A

Middle Resistance Hi point

Mid_res_hipt

60

Middle Resistance Lo point

Mid_res_lopt

2.1

High Resistance Range

Hi_res_rng

2002

60 V/5 A

15 A

High Resistance Hi point

Hi_res_hipt

200

High Resistance Lo point

Hi_res_lopt

24