Table 60501-1 Specifications and Supplemental Characteristics (continued)

Current Slew Rate:*

Rate #

30 A Range Step

3 A Range Step

Transition Time

1

0.5 A/ms

0.05 A/s

8.0 ms

2

1.2 A/ms

0.12 A/s

3.2 ms

3

2.5 A/ms

0.25 A/ms

1.6 ms

4

5 A/ms

0.5 A/ms

800

µ

s

5

12 A/ms

1.2 A/ms

320

µ

s

6

25 A/ms

2.5 A/ms

160

µ

s

7

0.05 A/

µ

s

5 A/ms

80

µ

s

8

0.12 A/

µ

s

12 A/ms

32

µ

s

9

0.25 A/

µ

s

25 A/ms

16

µ

s

10

0.5 A/

µ

s

0.05 A/

µ

s

12

µ

s

11

1.2 A/

µ

s

0.12 A/

µ

s

12

µ

s

12

2.5 A/

µ

s

0.25 A/

µ

s

12

µ

s

*AC performance specified from 3 to 60 V.

Voltage Slew Rate:

Rate #

Voltage Range Step

Transition Time*

1

1 V/ms

8.0 ms

2

2.5 V/ms

3.2 ms

3

5 V/ms

1.6 ms

4

10 V/ms

800

µ

s

5

25 V/ms

320

µ

s

6

50 V/ms

160

µ

s

7

0.1 V/

µ

s

85

µ

s

8

0.25 V/

µ

s

85

µ

S

9

0.5 V/

µ

s

85

µ

S

*Transition time based on low capacitance current source.

Resistance Slew Rate

(2

range): Uses the value programmed for voltage slew rate.

Resistance Slew Rate

(2 k and 10 k

ranges): Uses the value programmed for current slew rate.

Transient Current Overshoot (When programmed from 0A):

Range

Transient Current Level

Current Slew Rate

Overshoot*

30 A

3-30 A

All slew rates

0

1.5 A

0.5 A/

µ

s to 2.5 A/

µ

s

6%

1.5 A

0.5 A/ms to 0.25 A/

µ

s

0

3 A

3 A

All slew rates

0

1.5 A

0.13 A/

µ

s and 0.25 A/

µ

s

3%

1.5 A

0.05 A/ms and 0.05 A/

µ

s

0

*Overshoot may be higher during first five seconds of programming if unit has been operating at full current. Overshoot
values assume a total inductance of l

µ