background image

3

Table 60501-1 Specifications and Supplemental Characteristics (continued)

Transient Operation:

Continuous Mode

Frequency Range:

0.25 Hz to 10 kHz

Frequency Resolution:

4%

Frequency Accuracy:

3%

Duty Cycle Range:

3% to 97% (0.25 Hz to 1 kHz); 6% to 94% (1 kHz to 10 kHz)

Duty Cycle Resolution:

4%

Duty Cycle Accuracy:

6% of setting ± 2%

Pulsed Mode
Pulse Width:

50

µ

s ± 3% minimum; 4 s ± 3% maximum

Transient Current Level (0 to 3 A and 0 to 30 A ranges):

Resolution:

13 mA (3 A range); 130 mA (30 A range)

Accuracy:

± 0.1% ± 40 mA (3 A range); ± 0.1% ± 200 mA (30 A range)

Temperature Coefficient:

100 ppm/

°

C ± 5 mA/

°

C

Transient Resistance Level (0.067 to 2

, 2

to 2 k

, and 20

to 10 k

ranges):

Resolution:

8.6 m

(2

range); 2.1 mS (2 k

range); 0.2 mS (10 k

range)

Accuracy:

± 0.8% + 16 m

with > 3 A at input (2

range)

± 0. 3% + 5 mS with

6 V at input (2 k

range)

± 0.3% + 5 mS with

6 V at input (10 k

range)

Transient Voltage Level (0 to 60 V):

Resolution:

260

mV

Accuracy:

± 0.1% ± 300 mV

Temperature Coefficient:

150

ppm/

°

C ± 5 mV/

°

C

Current Readback:

Resolution:

9 mA (via GPIB); 10 mA (front panel)

Accuracy:

(after 30 minute wait): ± 0.06% ± 40 mA

Temperature Coefficient:

65

ppm/

°

C ± 3 mA/

°

C

Voltage Readback:

Resolution:

17 mV (via GPIB); 20 mV (front panel)

Accuracy:

± 0.05% ± 45 mV

Temperature Coefficient:

50 ppm/

°

C ± 1.2 mV/

°

C

Maximum Readback Capability:

65 to 70 V (typical)

Power Readback:

Accuracy:

±0.2% ± 2 W