Atec Agilent-60501B User Manual
Atec Equipment
2
Table 60501-1. Specification and Supplemental Characteristics
SPECIFICATIONS
DC Input Rating:
Current:
0 to 30 A
Voltage
: 3 to 60 V (minimum dc operation from 0 to 2 V for 0 to 30 A)
Power:
150 W at 40
°
C (derated to 112 W at 55
°
C)
A. OPERATING CHARACTERISTICS B. DERATED CURRENT DETAIL
Constant Current Mode:
Ranges:
0 to 3 A; and 0 to 30 A
Accuracy:
(after 30 second wait):
±
0.1% ± 40 mA (both ranges)
Resolution:
0.8 mA (3 A range); 8 mA (30 A range)
Regulation:
10 mA (both ranges)
Temperature Coefficient:
100 ppm/
°
C ± 3 mA/
°
C (both ranges)
Constant Resistance Mode:
Ranges:
0.067 to 2
Ω
; 2
Ω
to 2 k
Ω
; and 20
Ω
to 10 k
Ω
Accuracy:
± 0.8% ± 16 m
Ω
with
≥
6 A at input (2
Ω
range);
± 0.3% ± 5 mS with
≥
6 V at input (2 k and 10 k
Ω
ranges)
Resolution:
0.54 m
Ω
(2
Ω
range); 0.14 mS (2 k
Ω
range); 0.014 mS (10 k
Ω
range)
Regulation:
10 mV with remote sensing (2
Ω
range); 10 mA (2 k and 10 k
Ω
ranges)
Temperature Coefficient:
800 ppm/
°
C ± 0.8 m
Ω
/
°
C (2
Ω
range);
300ppm/
°
C ± 0.5 mS/
°
C (2 k and 10 k
Ω
ranges)
Constant Voltage Mode:
Range:
0 to 60 V
Accuracy:
± 0.1% ± 50 mV
Resolution:
16
mV
Regulation:
5 mV (remote sense); 40 mV (local sense)
Temperature Coefficient:
100 ppm/
°
C ± 5 mV/
°
C