Table 60501-1 Specifications and Supplemental Characteristics (continued)

Source Turn-On Current Overshoot:

Less than 10% of final value (in CC and CR modes when connected to power

supplies with voltage rise times of greater than 500

µ

s).

Programmable Short Circuit:

0.066

(0.04

typical)

Programmable Open Circuit:

20 k

(typical)

Drift Stability

(over an 8 hour interval):

Current:

± 0.03% ± 5 mA

Voltage:

± 0.01% ± 10 mV

Reverse Current Capacity:

50 A when unit is on; 20 A when unit is off

Weight:

3.2 kg (7 lbs.)

Table 60501-2. Programming Ranges

Function

Front Panel

Front Panel

HPSL Command

Range of Values

Key

Display

(Short Form)

Constant Current

Set Range

C:RNG value

"CURR:RANG value"

Low Range

0 and

3 A

High Range

> 3 A and

30 A

Set Main Level

CURR value

"CURR value"

Low Range

0 to 3 A

High Range

0 to 30 A

Set Slew Rate

(shift)

C:SLW value

"CURR:SLEW value"

Low Range

0.00005 to 0.25 (A/

µ

s)

High Range

0.0005 to 2.5 (A/

µ

s)

Set Transient Level

C:TLV value

"CURR:TLEV value"

same as main level

*Set Triggered Level

"CURR:TRIG value"

same as main level

Constant Resistance

Set Range

R:RNG value

"RES:RANG value"

Low Range

0 and

2

Middle Range

> 2

and

2 k

High Range

>2 k

and

10 k

Set Main Level

RES value

"RES value"

Low Range

0 to 2

Middle Range

2

to 2 k

High Range

20

to 10 k

Set Slew Rate

(shift)

Low Range

V:SLW value

"VOLT:SLEW value"

same as voltage slew

Middle/High Range

C:SLW value

"CURR:SLEW value"

same as current slew

Set Transient Level

R:TLV value

"RES:TLEV value"

same as main level

*Set Triggered Level

"RES:TRIG value"

same as main level

Constant Voltage

Set Main Level

VOLT value

"VOLT value"

0 to 60 V

Set Slew Rate

(shift)

V:SLW value

"VOLT:SLEW value"

0.001 to 0.5 (V/

µ

s)

Set Transient Level

V:TLV value

"VOLT:TLEV value"

same as main level

*Set Triggered Level

"VOLT:TRIG value"

same as main level