beautypg.com

Dc mode output, Ac + dc mode output, Ac input – Atec California-Instruments-15001iX User Manual

Page 6: System, Ix series - specifications, Remote control, Output relay, Output impedance, Power, Waveforms

background image

Output Relay

Push button controlled or bus
controlled output relay

Output impedance

Programmable Z on 3001

iX,

5001

iX and 15003iX for 50 Hz

fundamental

Resistive:

range

17 - 1000 m

resolution

4 m

accuracy

2 % FS

Inductive:

range

230 - 1000

µ

H

resolution

4

µ

H

accuracy

2 % FS

DC Mode Output

Maximum DC power at full scale
of DC voltage range:

Model:

Power

3001

iX

1500 W

5001

iX

2500 W

10001

iX

5000 W

15001

iX

7500 W

15003

iX

2500 W/ш 3ш
7500 W/ш 1ш

Voltage Ranges

User selectable voltage range
combinations:

Range:

High

Low

270 V

135 V

300 V

150 V

Load Regulation

see AC mode

Line Regulation

see AC mode

Output Noise

< 250 mV

rms

Typ

(20 kHz to 1 MHz)

< 500 mV

rms

Max

Max. DC Current Capability

Maximum DC current in lowest
DC range pair:

Model

270

range

135

range

3001

iX

5.65

11.1

5001

iX

9.25

18.5

10001

iX

18.5

37.0

15001

iX

27.75

55.5

15003

iX 1ш

27.75

55.5

9.25

18.5

Current Limit

Programmable

from 0 A to max. current for
selected range.

AC + DC Mode Output

Power

Full AC power if DC component
is less than 20 % of full scale
voltage. Full DC power if DC
component is above 20 %.

System

Non Volatile Memory storage

16 complete instrument setups
200 user defined waveforms

Waveforms

Waveform Types

Sine

Square

Clipped Sine, 0 - 20 % THD

User defined

User defined waveform storage

Four groups of 50 user defined
arbitrary waveforms of 1024
points for a total of 200. One
group can be active at a time.

Transient Programming

Transient Types

Voltage: drop, step, sag, surge,
sweep
Frequency: step, sag, surge,
sweep
Voltage and Frequency: step,
sweep

Transient List Parameters:

Voltage, Frequency, Time or
cycles, Slew rate, Waveform
shape, Phase angle, Repeat

Transient lists storage

up to 32 transient steps per list

Time resolution

1 msec

Time range

1 msec - 9999 sec

Maximum slew rate

50

µ

sec for 10% to 90% of full

scale change into resistive load

Waveform Acquisition

Channels

Voltage and Current for each
phase.

Memory Depth

4096 samples/channel.

Maximum Sample Rate

39.0625 Ks/s.

Triggering

Auto, Phase, Transient.

Trigger Delay

Pre-trigger

0 - 104 msec 1ø
0 - 312 msec 3ø

Post-trigger

0 - 1000 msec.

Display

Front panel Graphics Display
with cursors.

Bus Interface

Full bus access to waveform
acquisition system.

Remote Control

IEEE-488 Interface

IEEE-488 (GPIB) talker listener.
Subset:
AH1, C0, DC1, DT1, L3, PP0,
RL2, SH1, SR1, T6
IEEE-488.2 SCPI Syntax

RS232C Interface

9 pin D-shell connector
Handshake: CTS, RTS
Databits:

7,8

Stopbits:

1,2

Baud rate:

9600, 19200,
38400

IEEE-488.2 SCPI Syntax
Supplied with RS232C cable

System Interface

Inputs:

Remote shutdown
External Sync

Outputs:

Function Strobe

AC Input

Voltage

Model 3001iX:
187 - 264 V

AC

,(L-N, 1 Phase)

All other models:
Standard:

187 - 264 V

AC

,(L-L, 3 Phase)

Option -400:

360 - 528V

AC

,(L-L, 3 Phase)

(Input range must be specified
when ordering)

Current

Input Line Current (per phase)

Model:

187-
264V

360-
528V

3001

iX

30 A

N/A

5001

iX

24 A

12 A

10001

iX

48 A

24 A

15001

iX

72 A

36 A

15003

iX

72 A

36 A

Inrush Current per chassis

< 14 A rms. / 84 A

peak

for 200

µ

s

@ 187-264 V
< 8 A rms. / 36 A

peak

for 400

µ

s

@ 360-528 V

Line Frequency:

47 - 63 Hz

Efficiency:

75 % typical

Power Factor:

0.6 typical

Hold-up Time:

At least 10 ms

iX Series - Specifications

1

Note 1: Specifications are warranted over an ambient temperature range of 25°± 5° C.
Unless otherwise noted, specifications are per phase for a sinewave with a resistive
load and apply after a 30 minute warm-up period.

This manual is related to the following products: