Single-output: 200 w gpib, Continued), Specifications – Atec Agilent-6643A User Manual
Page 2

39
Single-Output: 200 W GPIB
(Continued)
Specifications
(at 0
˚
to 55
˚
C unless
otherwise specified)
6641A-
6643A-
6644A-
6645A-
6645A-
J04
J11
J09
J05
J06
Special Order
Special Order
Special Order
Special Order
Special Order
Option
Option
Option
Option
Option
Number of Outputs
1
1
1
1
1
GPIB
Yes
Yes
Yes
Yes
Yes
Output ratings
Output voltage
13 V
40 V
70 V
150 V
170 V
Output current (40
˚
C)
15.3 A
5 A
3 A
1.2 A
1 A
Maximum current (50
˚
C/55
˚
C)
13.77 A/13 A 4.5 A/4.25 A 2.7 A/2.55 A 1.08 A/1.02 A 0.9 A/0.85 A
Programming accuracy at 25
˚
C ±5
˚
C
Voltage
0.06% +
8.5 mV
17.5 mV
31 mV
65 mV
74 mV
Current
0.15% +
21 mA
6.7 mA
4.1 mA
1.7 mA
1.7 mA
Ripple and noise
from 20 Hz to 20 MHz
Voltage
rms
300 µV
450 µV
600 µV
900 µV
1 mV
peak-peak
3 mV
3.5 mV
6 mV
9 mV
10 mV
Current rms
8 mA
3 mA
1.5 mA
1 mA
1 mA
Readback accuracy at 25˚C ±5˚C
(percent of reading plus fixed)
Voltage
0.07% +
10 mV
30 mV
47 mV
100 mV
140 mV
+Current
0.15% +
15 mA
5 mA
3 mA
1.3 mA
1.3 mA
–Current
0.35% +
40 mA
12 mA
6.8 mA
2.9 mA
2.9 mA
Load regulation
Voltage
1 mV
3 mV
4.5 mV
7 mV
8 mV
Current
1 mA
0.25 mA
0.25 mA
0.25 mA
0.25 mA
Line regulation
Voltage
0.5 mV
1 mV
1.5 mV
2.5 mV
3 mV
Current
1 mA
0.25 mA
0.25 mA
0.25 mA
0.25 mA
Transient response time Less than 100 µs for the output voltage to recover to its previous level (within 0.1% of
the voltage rating of the supply or 20 mV, whichever is greater) following any step change in load current of up to
50% of rated current
Supplemental Characteristics
(Non-warranted characteristics determined by design and
useful in applying the product)
Average resolution
Voltage
3.5 mV
12 mV
1.4 mV
37.5 mV
42.5 mV
Current
5 mA
2 mA
1.2 mA
0.5 mA
0.5 mA
OVP
23 mV
62 mV
110 mV
250 mV
285 mV
OVP accuracy
260 mV
800 mV
1.5 mV
3 V
3.4 V
More detailed specifications at www.agilent.com/find/6640
High Performance DC Power Supplies speed and accuracy for test optimization
Application Notes:
10 Practical Tips You Need to
Know About Your Power Products
5965-8239E
10 Hints for Using Your Power Supply
to Decrease Test Time
5968-6359E
Understanding Linear Power Supply
Operation (AN1554)
5989-2291EN
Modern Connectivity -
Using USB and LAN I/O Converters
(AN 1475-1)
5989-0123EN
Supplemental Characteristics
for all model numbers
DC Floating Voltage: Output terminals
can be floated up to ±240 Vdc from
chassis ground
Remote Sensing: Up to half the rated
output voltage can be dropped in each
load lead. The drop in the load leads
subtracts from the voltage available
for the load.
Command Processing Time: Average time
required for the output voltage to begin to
change following receipt of digital data is
20 ms for the power supplies connected
directly to the GPIB
Output Programming Response Time: The
rise and fall time (10/90% and 90/10%) of
the output voltage is less than 15 ms. The
output voltage change settles within
1 LSB (0.025% x rated voltage) of final
value in less than 60 ms.
Down Programming: An active down
programmer sinks approximately
20% of the rated output current
Modulation: (Analog programming of
output voltage and current)
Input Signal: 0 to –5 V