Atec Agilent-3000-X Series User Manual
Page 22
22
Waveform measurements
Cursors**
• Single cursor accuracy:
±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
• Dual cursor accuracy:
±[DC vertical gain accuracy + 0.5% full scale]*
• Units: Seconds(s), Hz (1/s), Phase (degrees), Ratio (%)
Automatic measurements
Measurements continuously updated with statistics. Cursors track last selected
measurement. Select up to four measurements from the list below:
• Voltage: peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot,
average- N cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N
cycles, AC RMS- full screen (standard deviation), ratio (RMS1/RMS2)
• Time: period, frequency, counter, + width, - width, burst width, duty cycle, rise time, fall time,
delay, phase, X at min Y, X at Max Y
• Count: positive pulse count, negative pulse count, rising edge count, falling edge count
• Mixed: area- N cycles, area- full screen
Counter
Built-in frequency counter:
• Source: on any analog or digital channel
• Resolution: 5 digits
• Maximum frequency: bandwidth of scope
Waveform math
Arithmetic
f (g(t))
g(t): { add, subtract, multiply between any 2 channels}
f(t): {FFT(g(t)), differentiate d/dt g(t), integrate ∫ g(t) dt, square root √g(t) }
Enabled between any combination of two channels
Arithmetic
DSOX3ADVMath advanced waveform math option adds Ax + B, Square, Absolute,
Common Log, Natural Log, Exponential, Base 10 Exponential, LP Filter, HP Filter, Magnify,
Measurement Trend, Chart Logic Bus (Timing or State).
FFT
Up to 64 kpts resolution
Set FFT Window to: Hanning, Flat Top, Rectangular, Blackman-Harris
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics
* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
** 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div
sensitivity setting.
Display characteristics
Display
8.5-inch WVGA
Resolution
800 (H) x 480 (V) pixel format (screen area)
Graticules
8 vertical divisions by 10 horizontal divisions with intensity controls.
Format
YT, XY, and Roll
Maximum waveform update rate
> 1,000,000 wfms/s
Persistence
Off, infinite, variable persistence (100 ms - 60 s)
Intensity gradation
64 intensity levels