Atec Agilent-1670G User Manual
Page 7
7
Agilent Technologies 1670G Series
Specifications and Characteristics
Probes (general-purpose lead set)
Input resistance
100 k
Ω, ±2%
Parasitic tip capacitance
1.5 pF
Minimum voltage swing
500 mV, peak-to-peak
Threshold accuracy*
±(100 mV + 3% of threshold setting)
Maximum input voltage
±40 V peak
State Analysis
Minimum state clock pulse width
3.5 ns
Time tag resolution
[3]
8 ns or ± 0.1% (whichever is greater)
Maximum time count
between states
34.4 seconds
Maximum state tag
count between states
[3]
4.29 x 10
9
states
Minimum master-to-master
clock time*
6.67 ns
Minimum master-to-slave
clock time
0.0 ns
Minimum slave-to-master
clock time
4.0 ns
Clock qualifier
setup/hold
4.0/0 ns fixed
Timing Analysis
Sample period accuracy
0.01% of sample period
Channel-to-channel skew
2 ns typical (not > 3 ns)
Time interval accuracy
± (sample period accuracy + channel-to-channel
skew + 0.01% of time interval reading)
Minimum detectable glitch
3.5 ns
Triggering
Sequencer speed
>150 MHz
Maximum occurrence counter
1,048,575
Range width
32 bits each
Timer value range
400 ns to 500 seconds
Timer resolution
16 ns or 0.1% (whichever is greater)
Timer accuracy
±32 ns or ±0.1% (whichever is greater)
Operating Environment
Temperature
Instrument: 0°C to 55° C (+32°F to 131°F)
Disk media: 10°C to 40°C (+50°F to 104°F)
Probe lead sets and cables:
0°C to 65°C (+32°F to 149°F)
Humidity
Instrument: up to 95% relative humidity at +40° C
Disk media and hard drive: 8% to 85% relative
humidity
Altitude
4,572 m (15,000 ft)
[3]
Time or state tags halve the acquisition memory when there are no unassigned pods.
* Warranted Specifications
370 ohms
GROUND
1.5pF
7.4pF
100 K
ohm
Figure 15. Equivalent Probe Load for the
01650-61608 General-Purpose Lead Set.