Corrected system performance – Atec Agilent-E5063A User Manual
Page 3

3
The specifications in this section apply to measurements made with the Agilent E5063A network analyzer under the following
conditions:
• No averaging applied to data
• Environmental temperature of 23 °C (± 5 °C) with less than 1 °C deviation from the calibration temperature
• Response and isolation calibration performed
Corrected System Performance
System Dynamic Range
Description
Specification
Typical
System dynamic range at test port
1
(IF Bandwidth = 3 kHz)
100 kHz to 300 kHz
63 dB
300 kHz to 8.5 MHz
68 dB
8.5 to 100 MHz
91 dB
100 MHz to 4.34 GHz
92 dB
4.34 to 8.5 GHz
81 dB
8.5 to 13 GHz
75 dB
13 to 16 GHz
65 dB
16 to 18 GHz
62 dB
(IF Bandwidth = 10 Hz)
100 kHz to 300 kHz
88 dB
92 dB
300 kHz to 8.5 MHz
93 dB
97 dB
8.5 to 100 MHz
116 dB
122 dB
100 MHz to 4.34 GHz
117 dB
122 dB
4.34 to 8.5 GHz
106 dB
112 dB
8.5 to 13 GHz
100 dB
106 dB
13 to 16 GHz
90 dB
100 dB
16 to 18 GHz
87 dB
93 dB
1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The effective
dynamic range must take measurement uncertainty and interfering signals into account.