Test port output (source), Test port output frequency, Test port output power – Atec Agilent-E5063A User Manual
Page 11

11
Test Port Output (Source)
Test port output frequency
Description
Speciication
Typical
Frequency range
Option 245
Option 285
Option 2H5
100 kHz to 4.5 GHz
100 kHz to 8.5 GHz
100 kHz to 18 GHz
Resolution
1 Hz (100 kHz to 6.5 GHz)
2 Hz (6.5 to 13 GHz)
11 Hz (13 to 18 GHz)
Source stability
± 7 ppm (5 to 40 ºC)
CW accuracy
± 7 ppm
Test port output power
Description
Speciication
Typical
Nominal power (preset power)
–5 dBm
Range
100 kHz to 300 kHz
300 kHz to 8.5 GHz
8.5 to 18 GHz
–20 to –5 dBm
–20 to 0 dBm
–15 to –5 dBm
Resolution
0.05 dB
Level accuracy
At 50 MHz, –5 dBm, absolute
(level latness)
1
100 kHz to 300 kHz
300 kHz to 1 MHz
1 MHz to 4.34 GHz
4.34 to 8.5 GHz
8.5 to 12 GHz
12 to 18 GHz
± 0.9 dB
± 3.7 dB
± 2.0 dB
± 1.0 dB
± 1.6 dB
± 3.6 dB
± 5.8 dB
Level linearity
2
–10 to –5 dBm, 100 kHz to 300 kHz
–10 to 0 dBm, 300 kHz to 8.5 GHz
–10 to –5 dBm, 8.5 to 18 GHz
–20 to –10 dBm, 100 kHz to 8.5 GHz
–15 to –10 dBm, 8.5 to 18 GHz
± 1.6 dB
± 1.6 dB
± 1.8 dB
± 2.7 dB
± 2.9 dB
1. Level accuracy of other frequencies is taken at
–5 dBm, relative to 50 MHz reference unless otherwise stated. Level accuracy includes averaged total (non-)
harmonics power. Its transient factor is not included.
2. Level linearity given is relative to
–5 dBm unless otherwise stated. Level linearity includes averaged total (non-) harmonics power. The level accuracy needs
to be taken into account for test port output power level. Its transient factor is not included.